Chapter 5: Calibration and Verification
4. Specify the Calibration Type and calibration to use in each range. For each range, press the range
button and then select the calibration from the list. Only those calibrations specified by the
Calibration
Type
button will appear.
5. Press the
Save
As
button.
6. Enter a name for the calibration group and then press the
Save
button.
Preparing for Calibration
Save System Settings
Prior to performing any calibration procedure or system protocol, save the current system settings, in the event
they need to be re-loaded.
To save the system settings
1. Select the
Maintenance
tab.
2. Press the
Advanced
button.
3. Press the
System Properties
button.
4. Make sure that the USB flash memory drive is attached to the USB port
8
.
5. Press the
USB Backup
button.
N
OTE
:
The factory default values may be used; however, these values are not as accurate as the actual
calibrations.
If you want to use a factory default calibration for one of the ranges, do NOT press the default button and
then select the calibration for that range. The Analyzer will automatically include the factory default.
8.
One external and two internal USB ports are provided for your convenience. However, you can only export
data from one port at any time. The Analyzer will export to the first inserted USB flash storage device.
While this USB flash storage device is inserted, the Analyzer will not recognize any additional USB storage
devices inserted in other USB ports. The Analyzer can take up to 15 seconds to recognize the flash
storage device.