GE Healthcare
Senographe DS
Revision 1
Service Information and Procedures Class A 2385072-16-8EN
Job Card D/R A246 - Manual Compression Brake
Page no. 1301
Chapter 9
JC-DR-A-246.fm
Job Card D/R A246 - Manual Compression Brake
Chapter 9
1
SUPPLIES
2
TOOLS
•
•
Mammography compression scale with digital display
3
REQUIRED EFFORT
Personnel: 1 Field Engineer
Time: 30 minutes
4
SAFETY PRECAUTIONS
WARNING
The Gantry must be powered off.
5
PREREQUISITES
•
Lower the arm to obtain good access to
the Manual Compression Brake (1).
•
Switch off the Gantry power.
•
Remove the paddle.
•
Remove the compression covers (Refer to
Job Card PHY A044 - Remove/Reinstall
Gantry Covers
).
1