SUPPLIES
2
TOOLS
3
REQUIRED EFFORT
Personnel: 1 Field Engineer
Time: 30 minutes
4
SAFETY PRECAUTIONS
WARNING
The Gantry must be powered off.
5
PREREQUISITES
•
Switch off the Gantry power.
•
Remove the front covers, the column covers, and the bot-
tom front cover (2) (refer to
) for access to the
Gantry Readout (1).
6
PROCEDURE
6-1
Disassembly of the Gantry Readout
1. Disconnect the LCD connect cable (W204) from the Lift
Board (connector J15).
2. Carefully remove the LCD connect cable through the column.
3. Remove the four screws securing the Gantry Readout (screwdriver).
4. Remove the Gantry Readout.
6-2
Reassembly of the Gantry Readout
Reverse the disassembly procedure.