Page 44
g
5.3.3.6
Serial Port Configuration Parameters
The number of serial ports available for configuration (if any) is determined by the number
of USB to serial port adapters connected. A serial port requiring configuration is enabled
within the RSDetector unit for each device connected.
Table 13 Serial Ports Configuration Parameters
Parameter
Variable
Property Description
Choices
SerialPort
BaudRate
Determines the BAUD rate of the
serial port.
A drop down list
determines the
choices.
NumberOfBits
The number of bits used in the
transmission of a character.
7 or 8
Parity
Parity of the transmitted byte.
A drop down list
determines the
choices or Even,
Odd or None.
5.3.3.7
RSDetection Configuration Parameters
This selection covers a number of properties that govern the overall system operation.
Table 14 RSDetection Configuration
Parameter
Variable
Property Description
Choices
Control
BiasVoltageOnAt
Startup
If true, the high voltage bias
(nominally 400V) will be applied at
startup.
True
False
LoggingEnable
A global enable that when false will
disable all logging and if true logging
will be determined by the
measurement or event configuration.
True
False
NotifyEnable
A global enable that when false will
disable all notifications and if true
logging will be determined by the
measurement or event configuration.
True
False
Legacy
DateFormat
The older units allowed for 3 types of
date output. This corresponds to the
legacy DTF command.
MM_DD_YY
DD_MM_YY
YY_MM_DD
UnitAddress
The legacy commands used a single
character address for selecting which
unit was being addressed. This
corresponds to the UID legacy
command.
Single character
from ‘0’ to ‘Z’
excluding ‘@’