SECTION 3 – MAINTENANCE AND TROUBLESHOOTING

 

2022100-201 Rev C 

© 2010 General Electric Company 

Page 17 of 35

 

 

INTELLISENSE® Battery 

 

 

 
 
The Responder AED IntelliSense battery technology offers you the most advanced battery capabilities 
available for defibrillators. Responder AED IntelliSense batteries contain an integrated memory chip that   
automatically stores important usage information, enabling the battery to maintain a complete history of its 
operating life. The actual battery history can be reviewed using the RescueLink software. 
This history includes: 
 
 

 

 

 

• 

Battery Identification 

 

 

 

 

• 

Battery Type 

    • Original 

Date 

of Installation in an AED 

    • Number 

of 

Charges 

completed 

 

 

• 

Time in Operation (hours: minutes) 

    • Days 

of 

Standby 

Operation  

    • Battery 

Capacity 

Remaining 

 
 

BATTERY OPERATING LIFE 

The battery operating life depends on the type of battery

,

 actual usage and environmental factors.  

 
The following table represents the expected life of the Responder AED when used in Standby Mode. 
 
 

 

Model  

Estimated Shelf 
Life Warranty 

Typical 

Shocks 

2019437 (9142)

 

Lithium 

5 Years   

4 Years   

Up to 290 shocks 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

BATTERY SHELF LIFE 

The Responder AED batteries have a shelf-life of five years.  Shelf-life is defined as the length of time a 
battery can be stored, prior to installation into AED, without degrading its performance. 
 

Note:

 Storing the battery outside its specific range (0-50°C) will decrease battery life. 

 

Summary of Contents for Responder AED

Page 1: ...Service Manual 2022100 201rC...

Page 2: ...es Saving Lives SmartGauge STAR IntelliSense RescueReady RescueLink RHYTHMx and Survivalink are trademarks and registered trademarks of Cardiac Science Corp Responder is a registered trademark of Gene...

Page 3: ...tact your local GE customer service Please have the serial and model numbers available The serial and model numbers are located on the back of the Responder AED Responder AED is manufactured for GE Me...

Page 4: ...may be reproduced or transmitted in any form by any means without the express written permission of General Electric Company Information in this documentation is subject to change without notice Name...

Page 5: ...iv THIS PAGE INTENTIONALLY LEFT BLANK FOR YOUR NOTES...

Page 6: ...ING REQUIREMENTS 16 INTELLISENSE Battery 17 PADS 19 AED INDICATORS 20 SETTING THE AED INTERNAL CLOCK 22 SECTION 3 MAINTENANCE TROUBLESHOOTING 23 OVERVIEW 23 SELF TESTS 23 INDICATOR TROUBLESHOOTING TAB...

Page 7: ...vi THIS PAGE INTENTIONALLY LEFT BLANK FOR YOUR NOTES...

Page 8: ...er AED SAFETY TERMS AND DEFINITIONS The triangle attention symbol shown below left identifies the potential hazard categories The definition of each category is as follows DANGER This alert identifies...

Page 9: ...tempt to recharge the battery Any attempt to recharge the battery may result in an explosion or fire hazard WARNING Shock Hazard Do not disassemble the AED Failure to observe this warning can result i...

Page 10: ...g or moving of the patient may cause AEDs to improperly analyze the patient s cardiac rhythm Stop all motion or vibration before attempting a rescue CAUTION Serial Communication Cable The serial commu...

Page 11: ...with respect to electric shock fire and mechanical hazards only in accordance with UL 60601 1 CAN CSA C22 2 No 601 1 M90 EN60601 1 and EN60601 2 4 Conforms to UL Standard UL60601 1 Certified to CAN C...

Page 12: ...e Disposable Single patient use only Tear here to open Do not recharge battery Position of pads on the chest of patient Dispose of properly in accordance with all state province and country regulation...

Page 13: ...ronic equipment that must be recycled Manufacturer Authorized European Representative Fragile handle with care Keep away from rain Keep dry This way up Stacking limit by number General symbol for reco...

Page 14: ...gh the rescue using a combination of voice prompts audible alerts and visible indicators INDICATIONS FOR USE INTENDED USE The AED with STAR Biphasic Waveform is intended to be used by personnel who ha...

Page 15: ...ASYSTOLE THRESHOLD The asystole baseline to peak threshold is set at 0 08 mV ECG rhythms at or below 0 08 mV will be classified as Asystole and will not be shockable NOISE DETECTION The AED will detec...

Page 16: ...T and Normal Sinus Rhythms NSR The SVT Discriminator will only be applied to rhythms that fall between the Detection Rate and the SVT Rate The factory default setting for this feature is NO THERAPY FO...

Page 17: ...of 120 seconds can be modified from 60 180 seconds in ServiceLink OPERATOR TRAINING REQUIREMENTS Persons authorized to operate the AED must have all of the following minimum training Defibrillation tr...

Page 18: ...Type Original Date of Installation in an AED Number of Charges completed Time in Operation hours minutes Days of Standby Operation Battery Capacity Remaining BATTERY OPERATING LIFE The battery operat...

Page 19: ...Do not attempt to recharge the battery Any attempt to recharge the battery may result in an explosion or fire hazard CAUTION Lithium Sulfur Dioxide Battery Pressurized contents Never recharge short c...

Page 20: ...Used pads may cause patient to patient contamination ELECTRODE INSTALLATION 1 Open the lid of the AED 2 Place the package into the lid so that the expiration label is visible through the clear window...

Page 21: ...rder Integrity of the internal circuitry is good When the STATUS INDICATOR is RED maintenance is required Note When Status Indicator is RED or Service Indicator is illuminated device cannot be used to...

Page 22: ...sly repeats the Battery Low voice prompt To continue the rescue leave the lid open remove the battery and replace with a fresh battery If battery replacement takes longer than 60 seconds the first res...

Page 23: ...ote There is a 3 second delay between the time the AED lid is opened and the start of the rescue This 3 second delay is not included in the elapsed rescue time SETTING THE AED INTERNAL CLOCK The inter...

Page 24: ...es to RED Performs the self test If successful the STATUS INDICATOR reverts to GREEN Turns itself OFF if the lid is closed There are three types of automatic self tests The Daily Self Test checks the...

Page 25: ...a new battery The battery power is completely depleted Replace with a new battery If STATUS INDICATOR remains RED refer to the Responder AED for maintenance Call Customer Service or your local distrib...

Page 26: ...egrity 12 Close the lid Check the Integrity of the Service Indicator LED and Circuitry 1 Immediately after opening the AED lid press and hold the SHOCK button and confirm that the SERVICE LED is lit 2...

Page 27: ...ftware can be found in the ServiceLink manual provided on the ServiceLink software CD 1 Use ServiceLink to ensure the AED is set to a known protocol 1 Standard High Default value 200VE 300VE 300VE 2 S...

Page 28: ...nder AED 32 RHYTHMx ECG analysis performance 34 PARAMETERS OPERATION Semi Automatic shock advisory AUDIBLE ALERTS Voice Prompt Maintenance Alert VISIBLE INDICATORS Status Indicator Battery Status Indi...

Page 29: ...LITHIUM BATTERY SPECIFICATIONS Output voltage 12VDC max Batteries are non rechargeable Lithium contents 9 2g max Check local regulations for disposal information Model Estimated Shelf Life Warranty Ty...

Page 30: ...ned and manufactured to conform to the highest standards of safety and performance including electromagnetic compatibility EMC The Responder AED Model 2019198 and pads conform to the applicable requir...

Page 31: ...agnetic IEC 61000 4 8 2001 IEC 60601 2 4 2002 Section 36 202 8 AAMI DF39 Section 3 3 21 2 3 80A m 47 5Hz 1 320Hz ESD IEC 61000 4 2 Level 3 IEC 60601 2 4 2002 Section 36 202 2 6KV contact discharge 8KV...

Page 32: ...resistive load 2000 0 1500 0 1000 0 500 0 0 0 500 0 1000 0 1500 0 1 000 0 000 1 000 2 000 3 000 4 000 5 000 6 000 7 000 8 000 9 000 10 000 TIME ms The Biphasic Truncated Exponential BTE waveform uses...

Page 33: ...2 128 173 75 1430 5 8 973 3 2 116 156 100 1434 7 0 1007 3 2 108 146 125 1437 8 3 1027 3 2 102 138 150 1439 9 5 1040 3 2 98 132 175 1441 10 8 1049 3 2 95 128 Table A2 Low Variable Energy 200 VE Wavefor...

Page 34: ...The STAR Biphasic defibrillation waveform will deliver variable escalating energy that is customized to each patient s needs based upon a patient s thoracic impedance This customization adjusts for t...

Page 35: ...Energy Range 2 J Factory Default 1 200 126 260 2 300 170 351 3 300 170 351 Protocol 2 1 200 126 260 2 200 126 260 3 300 170 351 Protocol 3 1 150 95 196 2 200 126 260 3 200 126 260 Protocol 4 1 150 95...

Page 36: ...39 requirement and AHA recommendation of Sensitivitya of 90 Shockable Rhythm VT Meets AAMI DF 39 requirement and AHA recommendation of Sensitivity of 75 Non Shockable Rhythm NSR Meets AAMI DF 39 requi...

Page 37: ......

Reviews: