![GE ReadyToProcess WAVE 25 Operating Instructions Manual Download Page 78](http://html.mh-extra.com/html/ge/readytoprocess-wave-25/readytoprocess-wave-25_operating-instructions-manual_128043078.webp)
Equipment dimensions and
weight
The table below lists the outer dimensions and weights of the bioreactor system units.
Weight (kg)
Dimensions, W x D x H (mm)
System unit
24.0
404 x 560 x 205
Rocker
4.8
276 x 360 x 117
CBCU
3.8
275 x 280 x 115
Pump
4.5
475 x 430 x 60
Tray 10
7.3
740 x 480 x 70
Tray 20
9.5
800 x 610 x 70
Tray 50
1.7
475 x 430 x 230
Lid 10
3.3
740 x 480 x 245
Lid 20
3.9
800 x 610 x 260
Lid 50
Environmental conditions
The following general requirements must be fulfilled:
•
The room must have forced ventilation, adapted for use of CO
2
, O
2
and N
2
.
•
The system units should not be exposed to sources of heat, such as direct sunlight
The installation site must comply with the following specifications.
Environmental requirements
Requirement
Parameter
Indoor use only
Allowed location
15°C to 32°C
Ambient temperature, operation
20% to 80%, noncondensing
Relative humidity, operating
800 to 1060 mbar
Atmospheric pressure
Up to 2000 m
Altitude, operation
78
ReadyToProcess WAVE 25 Operating Instructions 29009597 AD
4 Installation
4.1 Site preparation
4.1.2 Site requirements