5-56
3-3-11
Auto Send Destination
Auto Send Destination appears in a drop down list with all available destinations.
When the user chooses “Send Images” or Closes a patient file when “Auto Send”
is checked, the images are transferred to the destination selected here. Available
destinations will be configured in the system configuration.
3-3-12
Images Tagged
Images Tagged can be set to default to All or None. When “All” is selected,
Images will automatically be tagged when entering the View/Film screen.
3-4Service User Interface / Effect on IUI
3-4-1
Dose Measurement
(Yes, No) If set to “no,” Total Dose and DAP (including their titles) do not appear
on IUI. If “yes” Total Dose and DAP (including their titles) appears on IUI.
3-4-2
Pulsed Fluoro
(Yes, No) If set to “no,” only Continuous Fluoro is available and selection for Flu-
oro rate is removed from IUI screens. If set to “yes” all options are available to the
user.
3-4-3
DSA Package
(Yes, No) If set to “no” DSA steps are not available for use or able to be set up in
Prot Edit.
3-4-4
DICOM Ethernet
(Yes, No) If set to “no” the system is not set up to send images to other stations,
Auto–send and Auto–send destinations are not available in user preferences.
3-4-5
DICOM Print
(Yes, No) If set to “no,” the system is not set up to print.
3-4-6
DICOM Worklist
(Yes, No) If set to “no” the tab for worklist does not appear. If set to “yes” the tab
for worklist is displayed.
3-4-7
Pediatric Option
(Yes, No) If set to "no", the system Pediatric mode is not available. The Default
Dose and Default Pediatric Grid buttons are not displayed in the User Prefer-
ences.
3-4-8
Measurement System
(English, Metric) Display all measurements according to the selection. This
includes FOV/Magnification.