5.3
Connections to the remote control socket
Introduction
Information such as motor speed and direction of rotation can be received from, and
sent to, other instruments. This feature can be used in a variety of situations, two of
which are described below.
Volume information to a fraction
collector
The stepper motor in the pump gives a pulse for every 0.4° of rotation. This angle corre-
sponds to a certain volume and the information can be used to determine fraction size
in a fraction collector. One pulse corresponds to approximately 1 mL when using tubing
with i.d. 3.1 mm in the pump.
For further details, see
Section 5.4 Calibration of volume information, on page 48
Connection to a controller
For use in completely automated systems, Peristaltic Pump P-1 can be controlled for
example from ÄKTA™ systems using I/O-signals.
For further information refer to the relevant controller’s Instruction Manual.
Peristaltic Pump P-1 Operating Instructions 28962201 AF
47
5 Operation
5.3 Connections to the remote control socket