A-10
Patient Data Module
2030047-001A
Appendix A–Technical Specifications
Cardiac output
Certifications
Method:
Thermodilution
Parameters Displayed:
Cardiac output, blood temperature, injectate temperature,
real-time cardiac output washout curve, last average CO
Cardiac output range:
0.2 to 15 liters per minute
Blood temperature range:
17°C to 42°C (62°F to 107°F)
Blood temperature
accuracy:
±0.5°C 17°C to 30°C
Injectate temperature
range:
0°C to 30°C (32°F to 86°F)
Injectate temperature
accuracy:
±0.3°C
Cardiac output review:
Accept/reject individual measurements and store average
Catheter sizes:
5, 6, 7, 7.5, or 8 French
Injuectate volume
selections:
3, 5, or 10
IEC/EN/UL 60601-1
CAN/CSA C22.2 No. 601.1
IEC/EN 60601-1-2
IEC/EN 60601-2-27
IEC/EN 60601-2-30
IEC/EN 60601-2-34
IEC/EN 60601-2-49
IEC/EN 60601-2-51
EN 12470-4
EN ISO 9919
ANSI/AAMI SP10
ANSI/AAMI EC11
ANSI/AAMI EC13
CE Marking: Medical Devices Directive - 93/42/EEC
JJG 760
YY 91079
Summary of Contents for PDM
Page 7: ...2030047 001A Patient Data Module 1 1 1 Introduction ...
Page 13: ...2030047 001A Patient Data Module 2 1 2 Equipment Overview ...
Page 21: ...2030047 001A Patient Data Module 3 1 3 Installation ...
Page 26: ...3 6 Patient Data Module 2030047 001A Installation ...
Page 27: ...2030047 001A Patient Data Module 4 1 4 Service Tool and Configuration ...
Page 38: ...4 12 Patient Data Module 2030047 001A Service Tool and Configuration ...
Page 39: ...2030047 001A Patient Data Module 5 1 5 Maintenance ...
Page 48: ...5 10 Patient Data Module 2030047 001A Maintenance ...
Page 49: ...2030047 001A Patient Data Module 6 1 6 Troubleshooting ...
Page 54: ...6 6 Patient Data Module 2030047 001A Troubleshooting ...
Page 55: ...2030047 001A Patient Data Module 7 1 7 Field Replaceable Units ...
Page 69: ...2030047 001A Patient Data Module A 1 A Appendix A Technical Specifications ...
Page 80: ...A 12 Patient Data Module 2030047 001A Appendix A Technical Specifications ...
Page 81: ...2030047 001A Patient Data Module B 1 B Appendix B Electromagnetic Compatibility ...
Page 90: ...B 10 Patient Data Module 2030047 001A Appendix B Electromagnetic Compatibility ...
Page 91: ......