Chapter 13: Advanced Applications
5495975-1EN Rev.9
13-8
© 2013-2017 General Electric Company. All rights reserved.
Repeat/Reject Analysis (RRA)
RRA is an option that allows classification and tracking of images that are considered unsuitable for
diagnostic purposes. The repeated and rejected images are retained on the system and can be exported
at any time. The system allows periodic generation of a statistical RRA report with Clinical Repeat Rate.
The Auto Tag function must be enabled (ON) in order to enable RRA on the system. Refer to
Preferences-Auto Tag (Quality Check) (p. 15-22)
for more information about enabling Auto Tag.
IMPORTANT!
The RRA tool is not linked to service requests; any system issues must be reported to ser-
vice.
Figure 13-5
Preferences - Repeat/Reject Analysis
Enable RRA
Follow this process to enable or disable RRA:
1. From the Worklist screen, click [UTILITIES].
2. Ensure Auto Tag is enabled on the system. Refer to
Chapter 15: Preferences-Auto Tag (Quality Check)
.
3. Select
Preferences > Repeat/Reject Analysis.
4. Click Enable RRA [ON] to enable the function.
Enable RRA [OFF] disables the function. RRA will be automatically disabled if Auto Tag is disabled.
5. Click [CLOSE].