Section 6.0 - Electrical Power On & Ground Checks
4.) Use a 0-750 AC voltmeter of 3/4% accuracy to measure the line-to-line voltages at L1, L2, &
L3.
-
Verify the highest line-to-line voltage does not exceed 1.02 times the lowest voltage.
-
Example:
If the lowest voltage equals 474, the highest voltage should not exceed 474 x
1.02 = 483.5 volts.
WARNING
THIS PROCEDURE MEASURES POTENTIALLY HAZARDOUS VOLTAGES. USE AND
FOLLOW LOCKOUT/TAGOUT PROCEDURES.
Record system voltages here:
Phase A: _______ Phase B: _______ Phase C: _________
6.7
System Power-Up
CAUTION
Verify all personnel have cleared the system before you turn on wall power.
1.) Turn ON the A1 breaker panel.
Note:
Do not stand in front of the main disconnect to turn on power.
2.) Turn ON all system power switches and breakers (PDU, gantry, table, console).
-
All PDU breakers
-
Make sure that the on/off button (on the front PDU panel) is ON for console power.
Signed
Date
TAG
&
LOCKOUT
No Load
Line to Line Voltages
Tap Connections
(All 3 phases must have same the
configuration)
Nominal
Maximum Range
(10%)
Phase A
Connection
Phase B
Connection
Phase C
Connection
480V*
432 to 528*
3-4*
3-4*
3-4*
460V
414 to 506
3-5
3-5
3-5
440V
396 to 484
3-6
3-6
3-6
420V
378 to 462
2-4
2-4
2-4
400V
360 to 440
2-5
2-5
2-5
380V
342 to 418
2-6
2-6
2-6
240V**
216 to 264**
1-4**
1-4**
1-4**
220V**
198 to 242**
1-5**
1-5**
1-5**
200V**
180 to 220**
1-6**
1-6**
1-6**
* Factory Default
** 2326492-3 PDU only
Table 4-1 PDU Line Tap Connections