Programming the VGA Optica
4-7
January 2006
Pressure Input
For more information on using these items, see
Process Pressure
on
Table 4-4: Pressure Input Options
Item
Function
Available Options
Input
Choose which pressure input channel is in use, or
disable this input.
V in (0-5), I in (4-20), or Use
Default
Units
Choose the units for the pressure input.
(See Table 4-1 on page 4-4 for
parameter chosen.)
Upper Limit
Set the pressure that corresponds to full-scale input.
(Enter a number.)
Lower Limit
Set the pressure that corresponds to zero volts or
4 mA input.
(Enter a number.)
Default
Pressure value to be used if the measured input is
disabled.
(Enter a number.)
Process
Enter a process pressure when it is different from the
pressure at the dew point sensor (see page 3-4).
Status
Set whether the process pressure is active (enabled).
Enable/Disable
Pressure
Enter the process pressure, if this feature is enabled.
(Enter a number.)
Summary of Contents for Optica
Page 1: ...GE Sensing Optica General Eastern Dew Point Analyzer Operator s Manual ...
Page 11: ...Chapter 1 ...
Page 20: ...Chapter 2 ...
Page 43: ...Chapter 3 ...
Page 58: ...Chapter 4 ...
Page 80: ...Chapter 5 ...
Page 94: ...Chapter 6 ...
Page 95: ...Network Based Programming Introduction 6 1 Programming Screens 6 1 ...
Page 99: ...Chapter 7 ...
Page 109: ...Appendix A ...
Page 114: ...Appendix B ...
Page 115: ...Humidity Equations and Conversion Chart Introduction B 1 Vapor Pressure B 1 Humidity B 2 ...
Page 119: ...Appendix C ...
Page 120: ...Configuring the Serial Interface Wiring to a Personal Computer C 1 ...
Page 122: ...Appendix D ...
Page 127: ...Appendix E ...
Page 128: ...Glossary ...
Page 130: ...Appendix F ...
Page 133: ...Appendix G ...