Hygrometer Calibration
The
Optica
unit can be sent to the National Institute of Standards and
Technology (NIST) in Gaithersburg, Maryland for certification or to
any National Standards lab for calibration against their primary
humidity standards. A calibrated instrument can then be used as a
transfer standard in local laboratories to calibrate lower echelon
instruments.
Caution!
Field calibration is not recommended.
Hygrometers used as calibration standards must have the following
characteristics:
•
The mirror thermometer must have suitable long-term accuracy
(such as that obtained with a platinum resistance thermometer).
•
A means should be provided for viewing the dew or frost
formation on the mirror.
Figure 1-2: Chilled-Mirror Hygrometer Diagram
41.2°F
Dew Point Temperature
(Precision Thermometer)
LED
Regulation
Thermoelectric
Heat Pump Power
Gain
Optical Reference
Summary of Contents for Optica
Page 1: ...GE Sensing Optica General Eastern Dew Point Analyzer Operator s Manual ...
Page 11: ...Chapter 1 ...
Page 20: ...Chapter 2 ...
Page 43: ...Chapter 3 ...
Page 58: ...Chapter 4 ...
Page 80: ...Chapter 5 ...
Page 94: ...Chapter 6 ...
Page 95: ...Network Based Programming Introduction 6 1 Programming Screens 6 1 ...
Page 99: ...Chapter 7 ...
Page 109: ...Appendix A ...
Page 114: ...Appendix B ...
Page 115: ...Humidity Equations and Conversion Chart Introduction B 1 Vapor Pressure B 1 Humidity B 2 ...
Page 119: ...Appendix C ...
Page 120: ...Configuring the Serial Interface Wiring to a Personal Computer C 1 ...
Page 122: ...Appendix D ...
Page 127: ...Appendix E ...
Page 128: ...Glossary ...
Page 130: ...Appendix F ...
Page 133: ...Appendix G ...