Humidity Equations and Conversion Chart
Humidity
Relative Humidity is defined as the ratio of the water vapor pressure
(e) to the saturation vapor pressure (e
S
) at the prevailing ambient or
dry bulb temperature (Ta):
(B-3)
Absolute humidity is expressed as water vapor density: water vapor
mass per unit volume of dry air, according to the following:
(B-4)
Water vapor content expressed as parts per million by volume is given
by the following:
(B-5)
Expressing water vapor content as parts per million by weight (or
mixing ratio) requires multiplication of the above by the ratio of the
molecular weight of water to that of air as given by the following:
(B-6)
See Figure B-1 on page B-3 for a graphical humidity conversion
chart.
%
RH
100 E
E
S
-----
⎝ ⎠
⎛ ⎞
100
E
W
T
D
(
)
E
WS
T
A
( )
--------------------
=
=
G
M
3
-------
216.7E T
D
(
)
T 273.16
+
--------------------------
=
PPM
V
10
6
E T
D
(
)
P E T
D
(
)
–
----------------------
=
PPM
W
0.622 10
6
×
E
P E
–
-----------
=
Summary of Contents for Optica
Page 1: ...GE Sensing Optica General Eastern Dew Point Analyzer Operator s Manual ...
Page 11: ...Chapter 1 ...
Page 20: ...Chapter 2 ...
Page 43: ...Chapter 3 ...
Page 58: ...Chapter 4 ...
Page 80: ...Chapter 5 ...
Page 94: ...Chapter 6 ...
Page 95: ...Network Based Programming Introduction 6 1 Programming Screens 6 1 ...
Page 99: ...Chapter 7 ...
Page 109: ...Appendix A ...
Page 114: ...Appendix B ...
Page 115: ...Humidity Equations and Conversion Chart Introduction B 1 Vapor Pressure B 1 Humidity B 2 ...
Page 119: ...Appendix C ...
Page 120: ...Configuring the Serial Interface Wiring to a Personal Computer C 1 ...
Page 122: ...Appendix D ...
Page 127: ...Appendix E ...
Page 128: ...Glossary ...
Page 130: ...Appendix F ...
Page 133: ...Appendix G ...