Maintenance
7-7
January 2006
Incorrect Dew Point
Display
If the dew/frost point reads incorrectly, first check the standard
preventive maintenance items:
1.
Clean the mirror (see
Cleaning the Sensor Mirror
on page 7-1).
2.
Balance the sensor optics (see
Balancing the Sensor Optics
on
An alternative method for checking the accuracy of the unit’s
electronics is to use a precision resistance decade box in place of the
platinum thermometer. Wire the decade box to the unit’s sensor
connector as shown in Figure 25, and verify that the resistance
settings shown in the table produce the temperatures shown.
Figure 7-2: Using a Resistance Decade Box
0.0
–10.0
–60.0
°C (±0.1)
Display Reading
60.0
10.0
Din Type 100 Ohms at 0°C 0.385 Ohms/°C
Resistance Table
RX10
123.24
RX0.01
RX0.1
RX1
76.33
100.00
103.90
96.09
Resistance
Ohms
3
Resistance
Decade Box
RX100
LOW
HIGH
14
4
15
140.0
–76.0
32.0
50.0
14.0
°F (±0.2)
Optica
Connector
Cable
Sensor
Optica
Connector
Cable
T
L
J
K
Sensor
D2
1123HK
Summary of Contents for Optica
Page 1: ...GE Sensing Optica General Eastern Dew Point Analyzer Operator s Manual ...
Page 11: ...Chapter 1 ...
Page 20: ...Chapter 2 ...
Page 43: ...Chapter 3 ...
Page 58: ...Chapter 4 ...
Page 80: ...Chapter 5 ...
Page 94: ...Chapter 6 ...
Page 95: ...Network Based Programming Introduction 6 1 Programming Screens 6 1 ...
Page 99: ...Chapter 7 ...
Page 109: ...Appendix A ...
Page 114: ...Appendix B ...
Page 115: ...Humidity Equations and Conversion Chart Introduction B 1 Vapor Pressure B 1 Humidity B 2 ...
Page 119: ...Appendix C ...
Page 120: ...Configuring the Serial Interface Wiring to a Personal Computer C 1 ...
Page 122: ...Appendix D ...
Page 127: ...Appendix E ...
Page 128: ...Glossary ...
Page 130: ...Appendix F ...
Page 133: ...Appendix G ...