NOVA

Brief Instruction Card

Operating Instructions

All functions and parameters of the NOVA can be set and recalled on the display at the feed table.

Display indicator for:
1  Developer temperature
2  Fixer temperature
3  Dryer step
4  OK signal
5  SET - key
6  +  key
7  -   key
8  Value display
9  Status indication / Menu step 

1
2
3

5

4

6

7

8

9

DEV
DRY

FIX

DEV
DRY

FIX

DEV
DRY

FIX

DEV
DRY

FIX

DEV
DRY

FIX

DEV
DRY

FIX

DEV
DRY

FIX

Setting the process parameters

-  Use the (+) and (-) key to scroll the program

-  Activate a process setting with the SET key, a dot will appear at the bottom richt side (9)

-  The parameter value can now be changed using the (+) or (-) key

-  After 10 sec or by pressing SET the changed value is saved  

and the display returns to the DEV temperature indication function 

Menu 

Process parameter 

Range  

Standard settings

Developer temperature 

22 - 37 °C 

28 °C

Fixer temperature 

22 - 37 °C 

28 °C

Dryer step 

1 - 6 

Step 4

Process setting 

1.5´ - 12´ 

Developer replenishment  

200 - 950 ml/m² 

900 ml/m²

Fixer replenishment 

500 - 1250 ml/m² 

1200 ml/m²

Water replenishment 

0 - 15 l/m² 

13 l/m²  *)

*) The setting of the water feed rate is an optional  feature which can only be made when using a Comfort Kit.

Developer temperature (°C)

Fixer temperature (°C)

Dryer step

Process setting

Developer replenishment (x10 ml/m²)

Fixer replenishment (x10 ml/m²)

Water feed rate (l/m²) 