![GE MP100 Series Service Manual Download Page 277](http://html1.mh-extra.com/html/ge/mp100-series/mp100-series_service-manual_4253514277.webp)
Preventive maintenance
2026419-033E
CIC Pro™
10-13
The system resource indicator should be checked periodically, and appropriate action
should be taken when needed.
Depending upon the color of the CIC Pro center system resource indicator, different
actions are required:
System resource restart procedure
When you are prompted to restart, always perform the safe shutdown procedure as
described.
1.
From the multi-patient viewer, select
Setup CIC > Service Password
.
2.
Type
mms_com
and then press
Enter
.
3.
At the
c:\Program Files\Marquette\CIC\<version#>
prompt, type
stop
and press
Enter.
Yellow
Warning! Available system resources are running low. A
Preventive Maintenance action is required.
Red
Warning! Available system resources are running low.
System restart is required to correct the problem. Patients
will not be monitored at this Central while the System is
restarting. If the system is not restarted now, it will restart
automatically in approximately XX minutes.
NOTE
Where “XX” represents the number in minutes ranging
from 0 to 60.
Icon color
Text
System resource indicator
System resource condition
Action required
Green
Normal - System resources are within
normal parameters
No service action is currently required.
Yellow (value equal or greater than 364)
Advisory - System has been running
continuously for at least 364 days without a
system resource reset. System resources
may be running low
Schedule a system resource restart.
Yellow (value less than 364)
Advisory - System resources are running
low
Contact GE Service.
Red (value equal or greater than 365)
Low - System has been running
continuously for at least 365 days without a
system resource reset. System resources
may be running low
Perform a system resource restart at the
earliest possible opportunity as part of
annual preventative maintenance.
Red (value less that 365)
Low - Available system resources are
running critically low.
Contact GE Service.
Any color other than green (value less than
10)
Low - Available system resources are
running critically low.
Perform a system resource restart and
contact GE Service immediately. Be
prepared to provide log files. For more
information, refer to
Summary of Contents for MP100 Series
Page 12: ...x CIC Pro 2026419 033E...
Page 13: ...2026419 033E CIC Pro 1 1 1 Introduction...
Page 27: ...2026419 033E CIC Pro 2 1 2 Equipment overview...
Page 53: ...2026419 033E CIC Pro 3 1 3 Licensing...
Page 60: ...3 8 CIC Pro 2026419 033E Licensing...
Page 61: ...2026419 033E CIC Pro 4 1 4 Service interfaces...
Page 74: ...4 14 CIC Pro 2026419 033E Service interfaces...
Page 75: ...2026419 033E CIC Pro 5 1 5 Installation...
Page 94: ...5 20 CIC Pro 2026419 033E Installation...
Page 95: ...2026419 033E CIC Pro 6 1 6 Configuration...
Page 187: ...2026419 033E CIC Pro 7 1 7 Checkout procedures...
Page 208: ...7 22 CIC Pro 2026419 033E Checkout procedures...
Page 209: ...2026419 033E CIC Pro 8 1 8 Troubleshooting...
Page 241: ...2026419 033E CIC Pro 9 1 9 Field replaceable units FRUs...
Page 243: ...Field replaceable units FRUs 2026419 033E CIC Pro 9 3 Hard drive Flash drive...
Page 244: ...9 4 CIC Pro 2026419 033E Field replaceable units FRUs Rack mount...
Page 245: ...Field replaceable units FRUs 2026419 033E CIC Pro 9 5 Mounting options...
Page 265: ...2026419 033E CIC Pro 10 1 10 Preventive maintenance...
Page 279: ...2026419 033E CIC Pro 11 1 11 Reload software...
Page 288: ...11 10 CIC Pro 2026419 033E Reload software...
Page 289: ...2026419 033E CIC Pro 12 1 12 Upgrade software...
Page 298: ...12 10 CIC Pro 2026419 033E Upgrade software...
Page 299: ...2026419 033E CIC Pro A 1 A Electromagnetic compatibility...
Page 305: ...2026419 033E CIC Pro B 1 B Backup and restore backed up data modules...
Page 309: ...Device compatibility 2026419 033E CIC Pro C 1 C Device compatibility...
Page 312: ...C 4 CIC Pro 2026419 033E Device compatibility...
Page 313: ......