Configuration
2026419-033E
CIC Pro™
6-49
5.
After making your selections, choose one of the following:
Click
OK
to apply your changes and close the
CIC Setup
window.
Click
Apply
to apply your changes without closing the
CIC Setup
window.
6.
Verify the new settings prior to monitoring patients.
Set the Display Configuration (non-mirror CIC Pro centers)
The
Display Configuration
tab allows you to format the CIC Pro center’s multi-
patient viewer with the required number of patient windows.
The amount of displayed patient data is dependent on the total number of patient
windows displayed in the multi-patient viewer. It is also dependent on the
Auto
Display Button
settings on the
Display Configuration
tab (e.g.,
Maximizing the
Waveform Length
or
Maximizing the Number of Waveforms
).
Modifications to
Display Configuration
are subject to licensing restrictions.
Licensing determines the number of beds displayed on the CIC Pro center screen.
Modifying the display to show more than the licensed number of beds will result in
blank slots in the display.
NOTE
You must remove admitted beds from the display before you can select a
Display
Configuration
that would eliminate those patient slots from the display. In the
multi-patient viewer, right-click on the patient bed you want to remove and
choose
Select Care Unit then Bed Number > None
.
Unit License Default: Full
Disclosure License Type
NOTE
If the default does not match the actual license, full disclosure does not work.
Display a list of the full disclosure license type.
Offline Storage
WARNING
POTENTIAL DATA LOSS—Do not allow a
NO COMM
(patient offline) event to exceed the time limit selected
in the
Offline Storage
setting. The patient’s full disclosure data is deleted if the time limit is exceeded.
Set the amount of time stored in
Offline Storage
. This exists as a precautionary measure when contact with
a monitor has been lost. This feature applies most often when the monitor is being used in the
Rover
application, since the monitor moves (
roves
) from room to room, on and off the network.
If a discharge or admit cycle occurs off the network for a bed with full disclosure data collection, the full
disclosure sub-system is unaware that the patient has changed. When the bed returns to the network, full
disclosure will present the data from the discharged and admitted patients as part of the more recent patient’s
data.
Offline Storage
helps to prevent this by assuming that the patient has been discharged if the monitor is
off the network for greater than the number of minutes selected with the
Offline Storage
control.
To modify the
Offline Storage
period, choose the time period from the drop-down list and select
Yes
to
confirm that you want to change the time period.
Start Data Storage
Designate how full disclosure is enabled for patients at the time of admission.
Bed List
Lists beds for which full disclosure data is automatically stored.
Restore
button
Clear any changes you made to the
Full Disclosure Defaults
settings and revert to the previous settings.
Option Function
Summary of Contents for MP100 Series
Page 12: ...x CIC Pro 2026419 033E...
Page 13: ...2026419 033E CIC Pro 1 1 1 Introduction...
Page 27: ...2026419 033E CIC Pro 2 1 2 Equipment overview...
Page 53: ...2026419 033E CIC Pro 3 1 3 Licensing...
Page 60: ...3 8 CIC Pro 2026419 033E Licensing...
Page 61: ...2026419 033E CIC Pro 4 1 4 Service interfaces...
Page 74: ...4 14 CIC Pro 2026419 033E Service interfaces...
Page 75: ...2026419 033E CIC Pro 5 1 5 Installation...
Page 94: ...5 20 CIC Pro 2026419 033E Installation...
Page 95: ...2026419 033E CIC Pro 6 1 6 Configuration...
Page 187: ...2026419 033E CIC Pro 7 1 7 Checkout procedures...
Page 208: ...7 22 CIC Pro 2026419 033E Checkout procedures...
Page 209: ...2026419 033E CIC Pro 8 1 8 Troubleshooting...
Page 241: ...2026419 033E CIC Pro 9 1 9 Field replaceable units FRUs...
Page 243: ...Field replaceable units FRUs 2026419 033E CIC Pro 9 3 Hard drive Flash drive...
Page 244: ...9 4 CIC Pro 2026419 033E Field replaceable units FRUs Rack mount...
Page 245: ...Field replaceable units FRUs 2026419 033E CIC Pro 9 5 Mounting options...
Page 265: ...2026419 033E CIC Pro 10 1 10 Preventive maintenance...
Page 279: ...2026419 033E CIC Pro 11 1 11 Reload software...
Page 288: ...11 10 CIC Pro 2026419 033E Reload software...
Page 289: ...2026419 033E CIC Pro 12 1 12 Upgrade software...
Page 298: ...12 10 CIC Pro 2026419 033E Upgrade software...
Page 299: ...2026419 033E CIC Pro A 1 A Electromagnetic compatibility...
Page 305: ...2026419 033E CIC Pro B 1 B Backup and restore backed up data modules...
Page 309: ...Device compatibility 2026419 033E CIC Pro C 1 C Device compatibility...
Page 312: ...C 4 CIC Pro 2026419 033E Device compatibility...
Page 313: ......