Advanced Programming
3-41
March 2008
Entering Moisture Curve
Data
If you refer to your Calibration Data Sheet you will notice that the
data points begin numbering at 0 (zero); therefore, the
MMS-3
displays the selected number of data points you entered (Step 4)
minus 1.
MH Curve Menu 1
Use the arrow keys to move the
brackets to
S/N
and press
[YES].
[S/N] #PTS CURVE
Probe Serial # 1
Enter the serial number and press
[YES]
. The probe serial number is
engraved on the probe mount.
Then press
[NO]
to return to the
Curve menu.
H: 123456
MH Curve Menu 1
Use the arrow keys to move the
brackets to
#PTS
and press
[YES].
S/N [#PTS] CURVE
Enter Total #PTS 1
Enter the total number of data
points (1 to 16) and press
[YES].
Refer to the Calibration Data
Sheet for the number of data
points.Then press
[NO]
to return
to the Curve menu.
14
MH Curve Menu 1
Use the arrow keys to move the
brackets to
CURVE
and press
[YES].
S/N #PTS [CURVE]
# DP/°C MH 1
If you want to begin entering
data at the data point displayed,
press the right arrow key to
move the bracket to DP/°C.
Otherwise, enter the desired data
point number. Press
[YES]
and
press the right arrow key.
12 +0.0 0.0000