Advanced Programming
3-11
March 2008
MMS-3 Error Codes
Modbus RS485
Connection
The
MMS-3
offers a standard RS232 port. It is necessary to supply an
RS232 - RS485 converter for the Modbus functionality.
Depending on the converter, the RS232 from the
MMS-3
is connected
to the terminal marked RS232 on the converter. The RS485 output
from the converter can be connected directly to the user's DCS system
(point to point) or to a multi-drop system
Auxiliary Input (cont.)
5145
0x1418 Chan 1, Aux2, mA
MSW
5146
0x1419 Chan 1, Aux2, mA
LSW
5147
0x141A Chan 1, Aux2 mA Error Code
0-32767
5148
0x141B Chan 2, Aux2, mA
MSW
5149
0x141C Chan 2, Aux2, mA
LSW
5150
0x141D Chan 2, Aux2 mA Error Code
0-32767
Table 3-3: MODBUS Register Map
Register Address
Measurement
Data
Table 3-4: Error Codes Displayed by MODBUS
No.
Error
No.
Error
No.
Error
No.
Error
0
No Error
9
No Data
18 Pressure Fault
27 User Div Zero
1
No Probe
10 No Cal
19 Aux1 Fault
28 User Math
2
Out of Range
11 Unknwn Probe
20 Aux2 Fault
29 User Missing Term
3
Over Range
12 Aux Fail
21 Ref Fault
30 User Extra Term
4
Under Range
13 Cal Error
22 Gnd Fault
31 User Missing Op
5
Not Available
14 No Channel
23 ADC Fault
32 User Extra Op
6
No Communication
15 O
2
Fault
24 Fluid Low!
33 User Too Complex
7
Error 10
16 Hygro Fault
25 User Re Ent
34 User Missing Left
Parentheses
8
No Reference
17 Temp Fault
26 User Invalid
35 User Missing Right
Parentheses