June 2003
2-2
Troubleshooting and Maintenance
Testing the Alarm
Relays
The Relay Test Menu enables you to either trip or reset the alarm
relays. To enter this menu use Table 2-1.
*The Settings Menu will prompt for a passcode (see
Entering the
Passcode
in Chapter 2 of the
Startup Guide
).
1.
At the Alarm Settings Menu, press the CHANNEL menu key to
scroll to the desired channel before pressing the TEST key.
2.
When you enter the Relay Test Menu, a screen similar to Figure 2-
1 appears. To trip or reset alarms, use the A
½¾
B menu key to
move the pointer to the desired alarm; then press the appropriate
menu keys. The word “TRIPPED” or “RESET” will appear in the
corresponding space.
Figure 2-1: Relay Test Menu
3.
To exit, press the DONE menu key until
Main Menu
appears on the
message line.
If the alarms will not trip or reset, make sure the alarms are set up
correctly as described in
Setting Up the Alarms
in Chapter 2,
Operation
, of the
Programming Guide
. Also, make sure the alarms
are wired correctly as described in
Connecting the Alarms
in Chapter 1, and that the alarm option is installed on the channel
card!
Table 2-1: Moving from Main Menu to Relay Test menu
Press the following menu keys:
To enter the:
SETTINGS*
Settings Menu
OUTPUTS
Output Settings Menu
ALARMS**
Alarm Settings Menu
TEST
Relay Test Menu
ALARM RELAY STATUS
Alarm A Alarm B
Reset Tripped
NOTE: Alarms SUSPENDED during test.
Relay Test Menu
HELP RESET A B DONE
½¾
Summary of Contents for Moisture Image 1 Series
Page 2: ...Moisture Image Series 1 Service Manual ...
Page 8: ...Chapter 1 ...
Page 31: ...Chapter 2 ...
Page 61: ...Appendix A ...