GE Analytical Instruments ©2006
7-2
DML 14291 Rev. A
Figure 7-1:
Assembly of Accurate NO Breath Kit.
Assembly of the Accurate NO Breath Kit
The breath kit is used for on-line exhaled NO measurements and must be
assembled before use. To assemble the breath kit:
•
Remove the Hans Rudolph
®
non-rebreathing valve from its package.
•
Unscrew the white mouthpiece from the non-rebreathing valve.
•
Install the white Sievers mouthpiece with two Luer ports by screwing the
mouthpiece into the valve clockwise until secure.
•
Unscrew the clear exhaust port fitting from the valve, making sure the spiral
diaphragm valve remains inside the non-rebreathing valve.
•
Install the Sievers exhaust port with a Luer adapter by screwing the port into
the valve until secure.