CALIBRATION: Accutracker Calibration
5-4
APEX Telemetry Transmitter
Revision D
401566-166
Accutracker Calibration
It is recommended that you follow the “Accutracker DX NIBP
Operational Tests” in Chapter 3, “Maintenance,” when you receive the
Accutracker and every 12 months thereafter. If the unit fails, return the
unit to GE Marquette Medical Systems Service and Supplies. See “How
to Reach Us” in Chapter 1.