background image

Publication Information

Publication Information

This document describes version 1.02 of the MAC VU360

 Resting ECG Analysis System, also referred to as the “product” or "system" or

"device". It does not apply to earlier product versions. Due to continuing product innovation, specifications in this document are subject to

change without notice.

12SL, CASE, CardioSoft, InSite ExC, MAC, MACCRA, MARS, MUSE, Marquette, MobileLink, and MULTI-LINK are trademarks owned by GE

Medical Systems 

Information Technologies

, Inc., a General Electric Company going to market as GE Healthcare. All other trademarks

contained herein are the property of their respective owners.

This product complies with the requirements concerning medical devices from the following regulatory bodies.

Date of first CE mark - 2018.

For more information about compliance, refer to the 

Regulatory and Safety Manual

 for this product.

The document part number and revision are on each page of the document. The revision identifies the document’s update level. The

revision history of this document is summarized in the following table.

Revision

Date

Comment

1

30 April 2020

Customer Release

2

28 February 2021

Update content for Software SP01

To access other GE Healthcare Diagnostic Cardiology documents, go to 

https://www.gehealthcare.com/en/support/support-documentation-

library

, and click 

Enter Customer Documentation Portal

.

To access Original Equipment Manufacturer (OEM) documents, go to the device manufacturer's website.

This document describes the MAC VU360

 Resting ECG Analysis System, also referred to as the” product”, “system”, or “device”. This

document is intended to be used by an operator of the MAC VU360

 Resting ECG Analysis system.

Installation should be performed by a qualified personnel.

The MAC VU360

 Resting ECG Analysis System is intended to be used under the direct supervision of a licensed healthcare practitioner, by

trained operators in a hospital or facility providing patient care.

This document provides information required for the proper use of the system. Familiarize yourself with this information and read and

understand all instructions before attempting to use this system. Keep this document with the Regulatory and Safety manual, and with the

equipment at all times, and periodically review it.

Illustrations in this document are provided as examples only. Depending on system configuration, screens in the document may differ from

the screens on your system. Patient names and data are fictitious. Any similarity to actual persons is coincidental.

Third-party Licenses

This product includes software developed by:

Linux Kernel organization (

https://www.kernel.org

)

NXP Semiconductors (

https://www.nxp.com

)

Apache Software Foundation (

http://www.apache.org

)

OpenSSL.org (

http://www.openssl.org

)

OpenSSH (

https://www.openssh.com/

)

GNU Foundation packages (

https://www.gnu.org

)

Gentoo software (

https://packages.gentoo.org

)

Boost Libraries (

http://www.boost.org

)

POCO Project (

https://pocoproject.org

)

2

  MAC VU360

 Resting ECG Analysis System

 2088531-370-2

Summary of Contents for MAC VU360

Page 1: ...Manual Version 1 02 2088531 370 2 Keep this manual with your Regulatory and Safety Manual and equipment at all times and periodically review it MAC VU360 Resting ECG Analysis System Operator s Manual...

Page 2: ...pment Manufacturer OEM documents go to the device manufacturer s website This document describes the MAC VU360 Resting ECG Analysis System also referred to as the product system or device This documen...

Page 3: ...ce to obtain the source code of the open source software used in the product if required Support GE Healthcare maintains a trained staff of application and technical experts to answer questions and to...

Page 4: ...the Device on Standby 27 Perform System Reset 27 3 Patient Information 29 Patient Information Screen Overview 29 Start a Test for a New Patient 31 Enter Patient Information 32 Update Patient Informati...

Page 5: ...ached to a Different Patient Test 81 Change the Order Attached to a Patient Test 82 Detach an Order from a Patient Test 82 Order Status 83 6 Work with the Files List 85 Review a Stored Patient Report...

Page 6: ...118 Storage 119 Other Cleaning and Disinfection Agents 119 Additional Information 120 12 Troubleshooting 121 System Errors 121 ECG Acquisition Errors 122 Printing Errors 123 Report Transmission Error...

Page 7: ...Glossary 167...

Page 8: ...CG measurement and interpretative analysis prints 12 leads or 15 leads of ECG and transmits ECG data to and from a central ECG cardiovascular information system Front and Rear View Item Part Name Desc...

Page 9: ...er 6 Front Bin Used for storage 7 Front Wheels and Lock Used to mobilize and lock the trolley in position The front wheels are fixed wheels and non conductive 8 Acquisition Module Cable Connects the A...

Page 10: ...e focus in a window or to select items from the screen 7 Shift Use to type a capital letter For example press Shift p to type a capital P 8 Space bar Enters a space in the text 9 Alt Switches between...

Page 11: ...AC Power LED Indicates various AC power states Green light indicates the device is plugged in and receiving power No light indicates the device is not plugged into AC power 15 Power on LED Indicates...

Page 12: ...d enable the Full Disclosure option 3 Date and Time Current local date and time in the configured date and time format To configure a date and time format refer to the MAC VU360 Resting ECG Analysis S...

Page 13: ...leges If you disable user authentication or configure with Technician ID access the Default user must to log on as a user with sufficient privileges to access a menu option 9 Start New Patient icon Se...

Page 14: ...during the recording of a rhythm NOTE A change to the filter speed or gain is applicable to the current patient For a new patient the values are reset to default settings 16 Start Rhythm icon Select...

Page 15: ...configured with Technician ID access the Default user is prompted to log on as a user with sufficient privileges 3 Service Displays the Service screen used to service the device Your administrator mus...

Page 16: ...ts The color of the icons change to show the level of battery life You may operate the device connected to the AC Mains power when the batteries are removed The device can also operate with a single b...

Page 17: ...is critically low and you should connect to AC power immediately No color but crossed out Battery Not Present AC Mains Power Shows that batteryB is not in the device and the AC Mains power is connecte...

Page 18: ...icon when connected to a LAN or WLAN network Table 4 LAN Icons Network Status Icon Status Description LAN Active The device is connected to a LAN LAN Connected The device is connected to a remote serv...

Page 19: ...bination works only for English keyboard To increase the brightness of the screen select Alt Arrow up keys on the keyboard To decrease the brightness of the screen select Alt Arrow down keys on the ke...

Page 20: ...ogged on to the device as the Default User If the user authentication mode is Technician ID enter the Technician ID to log on as a Default User Power Off the ECG Device 1 Before you stop the device co...

Page 21: ...ser Authentication Mode Description Full authentication with STAT login The Login screen displays with these fields when the device is powered on or unlocked The users below can log on to the device P...

Page 22: ...default password immediately after the first login Service user The username is Service This username is intended for use by Service personnel A user with the user management privilege can set the pa...

Page 23: ...e displays on the upper right corner of all of the screens you have access If your login fails see the following table Table 9 Login Errors Symptom Cause Solution You entered your username or password...

Page 24: ...a valid Technician ID to populate the Technician ID field 2 Select Continue You are logged on as the Default User The Acquisition screen displays You can perform tasks with Default User or Technician...

Page 25: ...Change Password The Change Password dialog box opens 2 Enter the current password and new password and confirm the new password 3 Select Change Password If the new password meets the password require...

Page 26: ...t L on the keyboard NOTE This only works when Full Authentication with STAT is enabled The device is locked Your username displays on the lock screen Unlock the Device 1 Perform one of the following s...

Page 27: ...procedure Unlock the Device on page 26 to unlock and log in to the device If user authentication is not configured you are returned to the last screen displayed when the device was put on standby mod...

Page 28: ...ges logs and user data will be lost and unrecoverable 2 Enter the serial number of the device in the Enter the system serial number field The Restore button enables when the serial number is entered c...

Page 29: ...if order management is disabled Read a patient s barcode with a barcode scanner Use a keyboard Do an Admission Discharge Transfer ADT query WARNING INACCURATE PATIENT DATA Patient data can be kept fro...

Page 30: ...s filled in from the Location ID configured in the System Settings 6 BPM Shows the real time beats per minute BPM The shown heart rate updates whenever the heart rate calculation algorithm reports a c...

Page 31: ...tion screen 1 Select the Start New Patient icon located in the upper right corner of the Acquisition screen If Then A patient test is open with unsaved data a message tells that unsaved patient data w...

Page 32: ...nformation You can scan a patient s barcode using an external barcode reader or the internal barcode reader of the Acquisition Module Before you use an external barcode reader make sure that The BRCD...

Page 33: ...the data scanned from the barcode and existing patient information a message displays Perform one of the steps as follows Select Use Scanned Data to populate the barcode data into the relevant fields...

Page 34: ...found on the device The orders show on the Orders tab in a filtered list Select the order you will to attach to the patient test One local order is found on the device The order fills in the Patient...

Page 35: ...mum of five records display on the Acquisition screen Select the ADT record to search the order using that selected record s Patient ID Only one ADT record is found on the MUSE system System searches...

Page 36: ...the order you will to attach to the patient test One local order is found on the device The order fills in the Patient Information screen if there is no mismatch with the patient data on the screen No...

Page 37: ...ID order found in that ADT record If multiple orders are found on the system The orders show on the Orders tab in a filtered list Select the order you will to attach to the patient test If only one o...

Page 38: ...hat no matching remote orders are found If remote order query is failed on the system A message displays in the notification area that remote order query has failed No ADT records are found on the MUS...

Page 39: ...uery is done only on Site 1 if the MUSE system version is MUSE v9 SP5 or earlier ADT query to the MUSE system for sites other than Site 1 requires MUSE v9 SP6 or later Enter or Edit Patient Informatio...

Page 40: ...your entries select Save The information is saved and the Patient Information screen closes Based on the Mandatory fields apply for Transmission or Acquisition settings the ECG report will not be acce...

Page 41: ...save time and prevent the necessary to restart the ECG The Hookup Advisor reports the status based on signals from each leadwire When you connect the RA R leadwire and another leadwire to the patient...

Page 42: ...e does not automatically expand when the overall Hookup Advisor status is yellow or red Only the Hookup Advisor lead quality status indicator and status messages display in the Notification Area 3 Lea...

Page 43: ...dicator and contains a checkmark The two circles to the left are black A message displays showing that the signal lead quality is good to acquire an ECG NOTE The background color of the Start ECG icon...

Page 44: ...isition you can accept the ECG If the quality of the ECG signal has problems during the 10 seconds ECG acquisition the status changes from green to yellow or red It is recommended to reject an ECG tha...

Page 45: ...to collapse it These changes apply only to the current patient test s If you start a new patient the changes reset to the values configured for the device Enable ACS Interpretation You can enable the...

Page 46: ...ACS algorithm is not executed Change Speed Gain and Filter The Acquisition screen displays the waveform based on the configured speed gain and filter After starting a new patient you can change the sp...

Page 47: ...ecord an ECG Manually on page 48 Automatically Acquire an ECG Start a test for a new patient For more information see Start a Test for a New Patient on page 31 Make sure that your administrator enable...

Page 48: ...stops The automatic ECG function is aborted and the device will function in manual ECG mode if An automatic ECG acquisition is stopped before 10 seconds of data is acquired The preview of an ECG that...

Page 49: ...of ECG data for analysis The Start ECG icon changes to Stop ECG and the 10 seconds count of acquisition progress is displayed on the icon You can cancel the acquisition before 10 seconds of data is re...

Page 50: ...smission queue until you enter the patient demographic data for the mandatory fields An error message displays in the notification area You need to complete the data for the mandatory fields and manua...

Page 51: ...enable the Full Disclosure option You can accept this preview to save the ECG patient report in the Files tab or reject it and start another ECG Table 16 ECG Tab Displaying a Preview of the ECG Patien...

Page 52: ...go back to the preview and accept or reject it 6 Full Disclosure tab Click the Full Disclosure tab from the patient record menu bar to navigate to the Full Disclosure waveform and application The ECG...

Page 53: ...ueue of pending reports to be sent to the configured automatic destination For more information see Display the Report Queue on page 99 NOTE Based on the Mandatory fields apply for Transmission settin...

Page 54: ...view an ECG Patient Report When you disconnect patient leads and there is no activity with the MAC VU360 device for 2 minutes all ECG reports will automatically close After the 10 seconds ECG is acqui...

Page 55: ...more information on the report formats and the standard layout of the ECG patient report see ECG Report Formats on page 130 3 Gain and Filter To change the waveform gain or filter select anywhere arou...

Page 56: ...nt on the MAC VU360 device To start an ECG for a new patient select Start New Patient For more information see Start a Test for a New Patient on page 31 To start a new 10 seconds ECG on the same patie...

Page 57: ...ing print jobs are cancelled Record a Rhythm Ensure that sufficient paper is available in the paper tray to print a rhythm report A rhythm report can be stored in digital form in the Files list or pri...

Page 58: ...ist The rhythm is not printed Go to step 4 if you want to stop recording the rhythm before its configured duration is reached otherwise go to step 5 Both The rhythm for the patient is digitally record...

Page 59: ...cription 1 Patient Information Banner Displays patient information such as patient first name last name gender age and so on Select anywhere in the Patient Information banner to edit patient informati...

Page 60: ...b Select the Live tab to return to the live waveform display in the Acquisition screen To start a new rhythm for the same patient select the Live tab to return to the live waveform display in the Acqu...

Page 61: ...ation with folder path PDF format Full Disclosure USB R W flash drive PDF format Full Disclosure DCP server destination MUSE v9 SP6 or higher or MUSE NX and MUSE DICOM Gateway Pro SP6 or higher PDF fo...

Page 62: ...ys on the notification bar in the lower right side of the screen in the format Destination_Name Job_Status For example if the destination name is USB and the job status is Failed the status displays a...

Page 63: ...u enter the patient demographic data for the mandatory fields If you manually transmit ECG report an error message Unable to transmit Incomplete patient data displays on the Acquisition screen You nee...

Page 64: ...ythm in the selected report format is sent to the printer The patient report is printed in the order in which it was received If no other patient reports are printing the report is printed immediately...

Page 65: ...DATA Incorrect patient information can cause patient data mismatch Data assigned to the wrong patient causes erroneous patient data that can affect diagnosis and treatment Be sure to check the patient...

Page 66: ...on message may display if you are trying to delete a patient report that has not yet been transmitted to the default destination if a message has been configured by your administrator Perform one of t...

Page 67: ...e Overview The Full Disclosure ECG option shows one lead of the patient waveform for a maximum of 5 minutes From this waveform you can create a Full Disclosure report FD Report or create a 12 lead or...

Page 68: ...creen displays the previous 5 minutes of Full Disclosure ECG data If you disconnect the acquisition module the ECG acquisition stops in 10 seconds 3 To change the lead click Select Lead All the config...

Page 69: ...atient Report see Accept or Reject an ECG Patient Report on page 50 d To review an ECG Patient Report see Review an ECG Patient Report on page 54 6 To generate a Full Disclosure report click FD Report...

Page 70: ...left and right arrows on the screen to see the next page When you record Full Disclosure ECG If the acquisition module is disconnected the Full Disclosure report displays empty space on the screen If...

Page 71: ...formation for the Full Disclosure report For more information see Edit Patient Information in a Patient Report on page 65 7 Copy icon Prints a copy of the Full Disclosure report For more information s...

Page 72: ...You can retrieve orders from an order management server such as a MUSE system that is connected to your network When the orders list is updated either automatically or manually new orders are populate...

Page 73: ...or Visit Number 4 Navigation arrows Navigates to the previous and next pages in the Orders list 5 Last Updated date and time Displays the date and time the order list was last updated 6 Refresh icon...

Page 74: ...Only one order can be associated with a patient test at any given time Orders cannot be attached to Transmitted patient reports or Digital Rhythm patient reports If you do not have permissions to edi...

Page 75: ...ually update the Orders list Manually Update the Orders List You can manually update the orders list at any time even if the auto update option is enabled 1 From the Acquisition screen select the Orde...

Page 76: ...SAP Pre Op Call back Routine If you select the Priority column again the sorting order is reversed You can change the sort order by selecting any one of the column headers The orders list is sorted in...

Page 77: ...ocation filter in the Orders list the filter icon changes to to indicate that the order list is filtered and does not display all orders If you select Then A pre configured filter group The orders lis...

Page 78: ...the device configured in the device settings 3 To display an expanded list of Orders select the Expand icon The Orders expanded list opens By default the applied filter is Current Patient Location and...

Page 79: ...ation and select Save to save the patient data 4 Record the ECG For more information see Record an ECG Manually on page 48 5 Verify that the status of the order in the Orders list is Attached Attach a...

Page 80: ...patient report All patient demographic fields included in the order are populated in the test overwriting the existing patient data The status of the order is changed to Attached 4 Update test demogr...

Page 81: ...nformation Using the Keyboard on page 39 Attach an Order that is Attached to a Different Patient Test NOTE Only one patient test can be associated with an order at any given time regardless of the sta...

Page 82: ...that can affect diagnosis and treatment Be sure to check the patient information screen for each new patient Make sure that you attach the correct order for the correct patient 1 Detach the incorrect...

Page 83: ...der to a patient test from the Orders list The order status changes from Open to Attached The MUSE server is notified to change the status of the corresponding order on the MUSE system from Open to Pe...

Page 84: ...Work with Orders NOTE The Attached order status is not included in the transmitted patient reports 84 MAC VU360 Resting ECG Analysis System 2088531 370 2...

Page 85: ...collapsed list Table 22 Files Collapsed List Item Name Description 1 Files tab Opens the Files collapsed list that stores the saved patient reports 2 Expand icon Opens the Files expanded list 3 Files...

Page 86: ...ient reports 2 Print List Prints the stored records from the Files list This button is enabled only if the stored records are available 3 Collapse icon Collapses the Files list 4 Files expanded list c...

Page 87: ...you can only view the patient reports you created in the current session 1 From the Acquisition screen select the Files tab The Files collapsed list opens 2 Select the Rhythm or Full Disclosure or EC...

Page 88: ...more information see Delete Stored Patient Reports from the Files List on page 95 5 Edit icon Select the Edit icon to expand the Patient Information screen and edit patient information for the patien...

Page 89: ...ing ECG Shared directory destination with folder path PDF Hilltop and Sapphire XML Sapphire XML format is available if you enable the XML format output in the Option Manager formats Digital Rhythm USB...

Page 90: ...To another configured destination 1 Select anywhere around the ellipsis icon on the lower left side of the screen to expand the Transmit menu 2 From the expanded Transmit menu select any configured d...

Page 91: ...on a message displays in the notification asking you to confirm if you want to re transmit the already transmitted reports Perform one of the actions below Select OK to re transmit the patient report...

Page 92: ...h you want to print a copy The selected patient report opens in a new tab An ECG patient report displays in the ECG tab A Rhythm report displays in the Rhythm tab A Full Disclosure report displays in...

Page 93: ...quisition settings the ECG report is not accepted transmitted or printed until you enter the patient demographic data for the mandatory fields If you manually print an error message Unable to print In...

Page 94: ...on the Acquisition Module Edit Patient Information in a Stored Patient Report Make sure that you have the privilege to open stored rhythm Full Disclosure or ECG patient reports from the Files list and...

Page 95: ...ns with a grey background indicating that this is a stored patient report 3 Edit the patient information using a keyboard See Enter or Edit Patient Information Using the Keyboard on page 39 4 Select S...

Page 96: ...e the patient reports The selected patient reports are deleted from the Files list An alert may be configured by your administrator to warn you before deletion of untransmitted reports If this alert i...

Page 97: ...transmitting to a designated location are temporarily stored in the Queue list The successfully transmitted digital rhythm FD report and ECG patient reports are immediately deleted from the Queue lis...

Page 98: ...e figure illustrates the Queue expanded list Table 26 Queue Expanded List Item Name Description 1 Queue tab Displays the list of patient reports in the transmission queue 2 Collapse icon Collapses the...

Page 99: ...lays the re tried transmission count in numbers If the device sends the report on the first attempt the Retry Count is 0 Job Failure Reason Displays the reason for the failed transmission If the devic...

Page 100: ...and then try again if necessary If the job is in Not Sent or Failed status a message displays asking you to confirm the deletion of the selected job 5 Select Delete to confirm the deletion The select...

Page 101: ...gins transmission as soon as the current patient report is transmitted If a patient report is successfully transmitted it is immediately deleted from the Queue Review the queue to confirm that the pat...

Page 102: ...ys in the Acquisition screen The list of recent patients is stored in the Patients list and you can view up to 500 recent patient records The following figure illustrates the Patients collapsed list T...

Page 103: ...psed list Patient ID Last Name First Name Gender Displays the following six columns in the expanded list Patient ID Last Name First Name DOB Gender Race By default the list is sorted by Last Name in a...

Page 104: ...tion see Start a Test for a New Patient on page 31 2 Select the Patients tab on the right side of the Acquisition screen The Patients collapsed list opens and displays a list of patients 3 Double tap...

Page 105: ...remote applications and you can open a maximum of three tabs at a time If the remote application is slow to load the following message displays Loading URL Any errors generated because of the link dis...

Page 106: ...105 2 Open the PDF To navigate the PDF do any of the following Move through the PDF Select to move to the next page in the PDF Select to move to the previous page in the PDF Enter a specific page numb...

Page 107: ...different paper size The paper tray is configured to use the appropriate paper size for the destination location Use the instructions to change the paper size by moving the spacers in the paper tray 1...

Page 108: ...ottom of the tray and the tabs on both sides of the spacer are affixed to the designated tab spaces on the tray 5 Push the paper tray back to the printer assembly Store Thermal Paper When imaged and s...

Page 109: ...opes or sheet separators made of polystyrene polypropylene or polyethylene do not degrade thermal traces However these materials afford no protection against fading from external causes Do NOT store t...

Page 110: ...me To make sure a fully charged battery in between acquisitions power off the system and connect it to an AC wall outlet until you use the system again This prolongs battery runtime The battery status...

Page 111: ...ect the system to an AC wall outlet 3 Charge the battery for 3 hours or until the battery status indicator displays a full charge Replace the Battery NOTICE BATTERY PACK DISPOSAL Do not dispose of the...

Page 112: ...ries into the battery compartment slots in the correct orientation and direction indicated by the arrow 4 Lift the battery compartment door in the upward direction and close it 112 MAC VU360 Resting E...

Page 113: ...tions that contain wax or wax compounds to clean the equipment Do not immerse the device in any liquid as this may corrode metal contacts and affect signal quality Do not drip or expose the writer ass...

Page 114: ...ce Manual Contact your local GE Healthcare service representative for further information For instruments produced by another manufacturer reference the manufacturer s instructions for use Visual Insp...

Page 115: ...or if you suspect a problem Check the pins that the leadwires plug into They should not be bent or loose Contact GE Healthcare Customer Support for any repairs needed Check the Acquisition Module plas...

Page 116: ...ion Module Ensure that thorough cleaning is performed to remove all visible soil and organic matter prior to disinfection See Cleaning the Acquisition Module on page 115 1 Dispense fresh Super Sani Cl...

Page 117: ...astor wheels are functioning properly If you notice any items that need repair contact an authorized service representative to make the repairs Discontinue using the device until the appropriate repai...

Page 118: ...ley on page 117 1 Dispense fresh Super Sani Cloth Wipe s from the canister 2 Apply disinfectant to the entire surface using fresh wipes Treated surfaces must remain visibly wet for a minimum of three...

Page 119: ...es is the recommended cleaning and disinfection solution which has been validated on the device However below products are compatible with the device and may be used for cleaning and disinfection It s...

Page 120: ...using appropriate equipment materials and personnel and achieves the desired result This requires validation and routine monitoring of the process Any deviation from the provided recommendations shoul...

Page 121: ...ist to create additional storage space Memory is full This ECG cannot be saved The device storage capacity is full Transmit patient reports to configured destinations and delete the transmitted report...

Page 122: ...lure The keyboard is not working Contact your GE Healthcare Service Support representative if the error persists Touchscreen failure The touchscreen is not working Contact your GE Healthcare Service S...

Page 123: ...If the error persists contact your GE Healthcare Service Support representative Printing Errors Table 29 Printer Errors Encountered During ECG Patient Report Printing Error Message Error Condition Ho...

Page 124: ...hm printing Table 31 Printing Errors Encountered During Stored Records List Printing Error Message Error Condition How to troubleshoot Printer error Paper jam detected Paper is jammed 1 Carefully remo...

Page 125: ...ror persists contact your GE Healthcare Service Support representative Printer error Restart system Contact Service Unknown error or hardware failure in printer Restart the system to check if the erro...

Page 126: ...um limit of 1 000 reports Wait for the reports in the queue to transmit and try again Unable to transmit Incomplete patient data Unable to transmit one or more reports Incomplete patient data One or m...

Page 127: ...ive Make sure that the USB flash drive is firmly inserted into the USB port Hilltop generation unsuccessful PDF generation unsuccessful Sapphire generation unsuccessful The file generation is unsucces...

Page 128: ...client certificate Invalid Client Private Key Password The client private key password is invalid Enter a valid client private key password Client Certificate is invalid The client certificate is inv...

Page 129: ...Incompatible firmware version of FIPS module Contact your GE Healthcare Service representative to replace the FIPS module with a compatible version FIPS initialization failed Reconfigure wireless If p...

Page 130: ...wo columns of six rows of waveforms Each column is 5 seconds wide and printed at 50 mm s One column is printed on each page This results in a two page report Multiple page report 2 by 5s 1 Rhythm Ld S...

Page 131: ...rt Formats Report Format Name Description One Page or Multiple Page Report Computer Graphic Record CGR Shows three columns of four rows of medians at 25 mm s on the left side and three rows of wavefor...

Page 132: ...econds wide and printed at 25 mm s The lower part isone row of 10 seconds of one lead The rhythm lead shown in the report is configurable One page report 5 by 2s 3 Rhythm Lds Shows two parts The upper...

Page 133: ...rt One page report Expanded Median Shows the following parts The upper part of the report is text including the measurements but not the interpretation and the measurement matrix which is the measurem...

Page 134: ...ntal frontal and sagittal planes There are four charts for each plane showing the entire cardiac cycle at the selected gain The gain can be selected from 20 mm mv 40 mm mV 80 mm V or 160 mm mV Each ch...

Page 135: ...ut the patient such as First Name and Last Name Age and or Date of Birth Gender Race Height and Weight Room Location 2 Vital Signs Displays information about the patient s vital signs such as Heart ra...

Page 136: ...CG patient report in the configured report format for the selected lead set 8 Pace Spikes Displays pace spikes for patients with a pacemaker The pace spikes represent pacemaker pulses NOTE When two pa...

Page 137: ...y Visit Number Medical History List of Medications Rhythm Report Format Rhythm reports contain patient information waveform data and ECG acquisition data A rhythm report is a continuous recording of a...

Page 138: ...tient If the report contains more than one page the BPM changes on each page of the waveform data 5 The waveform data The rhythm report contains waveform data for 12 leads configured for the rhythm re...

Page 139: ...11 The Gain of the ECG waveform measured in mm mV NOTE You can change the gain before or during the recording and or printing of a rhythm For the printed rhythm report the printing stops and restarts...

Page 140: ...25 mm s 1 second of data is shown in 25 mm 5 large squares on the rhythm report When recording 10 mm mV 1 mV of data is shown in 10 mm mV 2 large squares on the printout Full Disclosure FD Report Form...

Page 141: ...shown on the printed or transmitted FD report when the acquisition module is disconnected NOTE A square waves is shown on the printed or transmitted FD report when the lead is disconnected 5 Prints th...

Page 142: ...e lead that tells the gain on the printed FD report For the digital FD report the calibration pulse of the last gain selected during recording is shown on the full report 11 Prints the Speed of the EC...

Page 143: ...ree of excessive hair NOTE Do not use solvents to clean the skin solvents trapped under electrodes may lead to abnormal skin reactions 3 Apply the electrodes to the prepared sites WARNING ELECTRIC SHO...

Page 144: ...ng print jobs in the queue the Patient Information screen is expanded You can enter patient data for the new patient If there are pending print jobs from the previous patient test in the queue a messa...

Page 145: ...erior axillary line on the same horizontal level as 4 6 V6 violet C6 violet Mid axillary line on the same horizontal level as 4 and 5 7 LA black L yellow Left deltoid 8 LL red F green Above the left a...

Page 146: ...C6 violet Mid axillary line on the same horizontal level as 4 and 5 7 LA black L yellow Left deltoid 8 LL red F green Above the left ankle alternate placement upper leg as close to the torso as possib...

Page 147: ...al level as 4 and 5 7 LA black L yellow Left deltoid 8 LL red F green Above the left ankle alternate placement upper leg as close to the torso as possible 9 RL green N black Above the right ankle alte...

Page 148: ...ectrode Placement To acquire a Vector Loops ECG use the electrode placement shown in the following diagram Table 50 Vector Loops Electrode Placement Item AHA Label IEC Label Description 1 LA black L y...

Page 149: ...ine on the same horizontal level as 3 and 4 8 RA white R red Just below the clavicle of the right arm 9 H orange H light blue Back of the neck avoid the carotid artery and jugular vein 10 M orange M l...

Page 150: ...Fields Patient Information Text Box Names When an order is attached to a patient test all Patient Information text boxes are read only The text box names in the table with an asterisk are not 150 MAC...

Page 151: ...or 9 if the patient year of birth is between 2000 to existent year xx patient place of birth g patient gender male odd number female even number Patient ID Identification number given to the patient...

Page 152: ...or in the long formats as follows yyyymmdd xxgc yyyymmdd xxgc yyyymmddxxgc yy and yyyy patient year of birth dd patient day of birth mm patient month of birth or patient age NOTE If or is not before t...

Page 153: ...of birth is between 2000 to current year g patient gender male odd number female even number xx patient place of birth Mandatory fields apply for The mandatory fields that can be configured for Transm...

Page 154: ...o 232 cm Weight Patient weight in pounds lb or kilograms kg refer to the configured unit of measurement Maximum 3 characters 0 to 999 lb 0 to 454 kg Gender Patient gender Not Applicable Male Female Da...

Page 155: ...c American Indian Eskimo Hawaiian Pacific Islander Asian Unknown Other Order Number Unique order number given to a patient test If the order number is given by the computer when an order is attached t...

Page 156: ...acters are supported Bed Number The bed number where the ECG is to be done Maximum 32 characters A to Z a to z 0 to 9 All characters are supported Priority Priority of ECG patient test Not Applicable...

Page 157: ...All characters are supported Referring MD ID Referring physician ID Maximum 5 characters 0 to 65534 Technician Technician doing the ECG If you are a local user and your user account is configured wit...

Page 158: ...o z 0 to 9 All characters are supported Number 10 characters Extra optional one special character or in the beginning 0 to 9 Yes or No or Unknown Question 1 Question 2 Question 3 Question 4 The name o...

Page 159: ...Values Attending MD First Name Attending physician first name Maximum 20 characters A to Z a to z 0 to 9 All characters are supported Attending MD ID Attending physician ID Maximum 5 characters 0 to 6...

Page 160: ...m to 62 in 158 cm Width 19 in 48 cm Depth 28 in 71 cm System with fixed height premium trolley Handle height 35 in 89 cm or 37 in 94 cm Maximum height 56 in 142 cm to 58 in 147 cm Width 19 in 48 cm De...

Page 161: ...le 54 Keypad Specifications Item Description Type Sealed elastomer membrane keyboard with tactile feedback Keys Function keys Alphanumeric keys Writer controls Trimpad cursor controls Keypad Keypad an...

Page 162: ...rge when device is off standby Battery capacity The system shall be able to continuously display scrolling waveforms without printing for a minimum of six hours when the maximum total battery capacity...

Page 163: ...g Specifications Item Description ECG Interpretation Marquette 12SL ECG Analysis Program for Adults and Pediatrics Computerized measurements 15 lead analysis includes measurements of user selectable a...

Page 164: ...ications Item Description Operating Conditions Temperature 10 C to 40 C 50 to 104 F Relative Humidity RH 20 to 95 non condensing Atmospheric Pressure 70 to 106kPa Transport Storage Conditions Temperat...

Page 165: ...ype CF defibrillation proof applied part Network Specifications Table 63 Network Specifications Item Description Frequency bands of transmission 2 401 2 461 GHz 5 180 5 825 GHz Maximum radiated power...

Page 166: ...non FIPS PSK PEAP MSCHAPV2 PEAP GTC EAP TLS TTLS MSCHAPV2 TTLS GTC Authentication Methods FIPS PSK PEAP GTC EAP TLS Wired Specifications Ethernet interface 802 3 Ethernet interface via RJ45 connector...

Page 167: ...e amplitude of the waveforms A higher gain makes the amplitude of the waveform appear higher a lower gain makes the amplitude of the waveform appear lower The 10 5 mm mV setting is used to display the...

Page 168: ...53223 USA 1 414 355 5000 1 800 558 7044 US Only GE Medical Systems Information Technologies GmbH Munzinger Stra e 5 79111 Freiburg Germany 49 761 45 43 0 www gehealthcare com GE Medical Systems Infor...

Reviews: