Revision E
MAC
5500 resting ECG analysis system
4-19
2020299-020
Troubleshooting: System Errors
System Errors
The following errors may occur while you are operating this system. You may be
required to perform some action.
If you perform the recommended actions and the condition still remains, contact
authorized service personnel. See “How to Reach Us” to find out how to contact GE
Medical Systems
Information Technologies
.
NOTE
For information about troubleshooting the MobileLink Standard Security
option, see “MobileLink Installation & Troubleshooting Guide”
(PN 2020299-021).
For information about troubleshooting the MobileLink Ultra High Security
option, see “MobileLink UHS Installation & Troubleshooting Guide”
(PN 2020300-051).
Problem
Cause
Solution
appears on the screen.
No battery is installed in the system.
Install a battery and connect the system to
an AC wall outlet to charge the battery.
flashes intermittently.
The battery charge is low.
Connect the system to an AC wall outlet to
charge the battery.
appears on the screen.
The writer door is open.
Close the writer door.
The system does not power up when
operating from battery power.
The battery is empty.
Connect the system to an AC wall outlet to
charge the battery.
The system shuts down when operating
from battery power.
Battery is empty, or the Automatic Shutdown
feature is enabled.
Connect the system to an AC wall outlet to
charge the battery, or power on the system.
“X” Lead disconnected message appears. Electrode(s) disconnected.
Reconnect the electrode(s).
MODEM ERROR. The remote device is
not responding. Would you like to retry?
Modem not connected. (If wireless option,
client bridge not connected.)
Connect and retry.
(Wireless option only) MAC 5500 is not
within range of access point.
Relocate MAC 5500 to within range of
access point and retry transmission.
Cannot use the system because Device
Password does not work.
Device Password has been changed or has
not been adequately communicated to the
staff.
Override the Device Password prompt by
pressing the following keys at the same time:
63A, 64A, 65A
i
Summary of Contents for MAC 5500
Page 1: ...g MAC 5500 Resting ECG Analysis System Service Manual 2020299 020 Revision E GE Healthcare ...
Page 9: ...Revision E MAC 5500 resting ECG analysis system 1 1 2020299 020 Introduction 1 Introduction ...
Page 10: ...1 2 MAC 5500 resting ECG analysis system Revision E 2020299 020 Introduction For your notes ...
Page 55: ...Revision E MAC 5500 resting ECG analysis system 3 1 2020299 020 Installation 3 Installation ...
Page 56: ...3 2 MAC 5500 resting ECG analysis system Revision E 2020299 020 Installation For your notes ...
Page 105: ...Revision E MAC 5500 resting ECG analysis system 5 1 2020299 020 Maintenance 5 Maintenance ...
Page 106: ...5 2 MAC 5500 resting ECG analysis system Revision E 2020299 020 Maintenance For your notes ...
Page 137: ...Revision E MAC 5500 resting ECG analysis system 6 1 2020299 020 Parts List 6 Parts List ...
Page 138: ...6 2 MAC 5500 resting ECG analysis system Revision E 2020299 020 Parts List For your notes ...
Page 175: ...Revision E MAC 5500 resting ECG analysis system Index 1 2020299 020 Index ...
Page 176: ...Index 2 MAC 5500 resting ECG analysis system Revision E 2020299 020 ...
Page 178: ...Index Index 4 MAC 5500 resting ECG analysis system Revision E 2020299 020 ...
Page 179: ......