Troubleshooting: Diagnostic Tests
External Modem Test
Connect a modem to
COM 2
and select
External
Modem
Test
. The test returns the
modem ID number, firmware rev, and current parameter settings. If communication
with the modem is unsuccessful, the ID and firmware rev display
N/A
.
Internal Modem Test
Select
Internal
Modem
Test
. The
Internal Modem Interrogation
screen displays. The
test returns the modem ID number, firmware rev, and current parameter settings. If
communication with the modem is unsuccessful, the ID and firmware rev display
N/
A
. Press any key to exit this test.
Ethernet Module Test
Select
Ethernet Module Test
. The
Ethernet Module Interrogation
screen displays.
The test returns the device IP address and the subnet mask information. If the
Ethernet module test is unsuccessful, the ID and firmware rev display
N/A
. Press any
key to exit this test.
Acquisition Module Test
Follow the instructions on screen.
NOTE
A shorting bar is required to perform this test.
Tests if the acquisition module is powered
Tests if the acquisition module is communicating
Displays the acquisition module lead wire noise
Indicates when one of the three acquisition module buttons is pressed
Displays software version of acquisition module
Analog I/O Tests
Analog Output Test
Follow the instructions on screen to monitor the analog outputs using an
oscilloscope. The outputs monitored are:
+12V
DC Output 1
DC Output 2
ECG Output
TTL Trigger Output
Four sets of outputs are possible. Select the output sets using the arrow pad.
For example, to test DCOut 1, connect oscilloscope to pin 2 and ground to pin 4 or 5
and select DCOut 2 0V and check oscilloscope for reading of 0 volts. repeat for
other selections. If any of these readings do not match, CPU may need to be
replaced.
Summary of Contents for MAC 5500
Page 1: ...g MAC 5500 Resting ECG Analysis System Service Manual 2020299 020 Revision E GE Healthcare ...
Page 9: ...Revision E MAC 5500 resting ECG analysis system 1 1 2020299 020 Introduction 1 Introduction ...
Page 10: ...1 2 MAC 5500 resting ECG analysis system Revision E 2020299 020 Introduction For your notes ...
Page 55: ...Revision E MAC 5500 resting ECG analysis system 3 1 2020299 020 Installation 3 Installation ...
Page 56: ...3 2 MAC 5500 resting ECG analysis system Revision E 2020299 020 Installation For your notes ...
Page 105: ...Revision E MAC 5500 resting ECG analysis system 5 1 2020299 020 Maintenance 5 Maintenance ...
Page 106: ...5 2 MAC 5500 resting ECG analysis system Revision E 2020299 020 Maintenance For your notes ...
Page 137: ...Revision E MAC 5500 resting ECG analysis system 6 1 2020299 020 Parts List 6 Parts List ...
Page 138: ...6 2 MAC 5500 resting ECG analysis system Revision E 2020299 020 Parts List For your notes ...
Page 175: ...Revision E MAC 5500 resting ECG analysis system Index 1 2020299 020 Index ...
Page 176: ...Index 2 MAC 5500 resting ECG analysis system Revision E 2020299 020 ...
Page 178: ...Index Index 4 MAC 5500 resting ECG analysis system Revision E 2020299 020 ...
Page 179: ......