Revision E
MAC
5500 resting ECG analysis system
5-19
2020299-020
Maintenance: Disassembly Guidelines
Printhead Replacement
Removal
1.
Remove the top cover following the procedure above.
2.
Using a Phillips head screw driver, remove the two screws that hold the
printhead to the printhead mounting plate.
3.
Open the writer assembly, disconnect and remove the printhead.
Reassembly
1.
Record the resistance value of the new printhead.
2.
Connect the new printhead to the ribbon cable.
3.
Hold the new printhead FIRMLY in place against the two metal tabs on the
printhead mounting plate, then tighten the two screws.
4.
Replace the top cover and power up the unit.
5.
Go to the Setup menu and enter the new printhead resistance value.
6.
Run a Writer Test test (See Chapter 5).
COMM Board Replacement
Use the following procedure to replace the communications board.
1.
Remove the AC power cable and battery.
2.
Using a Torx #10 screwdriver, remove the two screws from the panel
surrounding the LAN and modem ports.
3.
Grasp the side edges of the communications board and work it back and forth in
the slot as you pull it out along its rails.
Summary of Contents for MAC 5500
Page 1: ...g MAC 5500 Resting ECG Analysis System Service Manual 2020299 020 Revision E GE Healthcare ...
Page 9: ...Revision E MAC 5500 resting ECG analysis system 1 1 2020299 020 Introduction 1 Introduction ...
Page 10: ...1 2 MAC 5500 resting ECG analysis system Revision E 2020299 020 Introduction For your notes ...
Page 55: ...Revision E MAC 5500 resting ECG analysis system 3 1 2020299 020 Installation 3 Installation ...
Page 56: ...3 2 MAC 5500 resting ECG analysis system Revision E 2020299 020 Installation For your notes ...
Page 105: ...Revision E MAC 5500 resting ECG analysis system 5 1 2020299 020 Maintenance 5 Maintenance ...
Page 106: ...5 2 MAC 5500 resting ECG analysis system Revision E 2020299 020 Maintenance For your notes ...
Page 137: ...Revision E MAC 5500 resting ECG analysis system 6 1 2020299 020 Parts List 6 Parts List ...
Page 138: ...6 2 MAC 5500 resting ECG analysis system Revision E 2020299 020 Parts List For your notes ...
Page 175: ...Revision E MAC 5500 resting ECG analysis system Index 1 2020299 020 Index ...
Page 176: ...Index 2 MAC 5500 resting ECG analysis system Revision E 2020299 020 ...
Page 178: ...Index Index 4 MAC 5500 resting ECG analysis system Revision E 2020299 020 ...
Page 179: ......