Revision B
MAC
™
5000 resting ECG analysis system
5-7
2024917-010
Troubleshooting:
Diagnostic Tests
Diagnostic Tests
Introduction
Verify that the MAC 5000 resting ECG analysis system operates
properly by running the diagnostic tests. These tests check the operation
of the display screen, speaker, keyboard, thermal writer, battery, and
communication. Detailed information displays on screen.
Loading the System Diagnostics
1. Select
Main Menu
on the Resting screen.
2. Select
More
.
3. Select
System
Setup
.
4. At the prompt type the word “system”, the password set at the
factory, then press the
Enter
key. If the password was not changed,
the
System Setup
menu appears. If the menu does not appear, use
the master password. If the system’s unique password is inaccessible,
create one following the instructions in “Substitute Master
Password” later in this section.
5. When the
System Setup
menu displays, hold down
Shift
and press
F5
(
Shift
+
F5
).
6. Type
prod
at the service password prompt.
7. The
System Diagnostics
menu appears.
Substitute Master Password
If you do not have access to the system’s password, you can create a
master password as follows.
1. At the prompt for the system password, enter
meimac
. A random
6-digit number displays on the screen. For example, 876743.
2. Write the number down and create a new 6-digit number by adding
alternating digits from the random number as follows. Add:
first and third digits,
second and fourth digits,
third and fifth digits,
fourth and sixth digits,
fifth and first digits, and
sixth and second digits.
Disregard the 10s column when adding the digits. The new number
from the example above would be 440020.
3. Enter the new number, then press the
Enter
key. The
System Setup
menu displays. This process only works once, so you should
reprogram the password permanently.
Summary of Contents for MAC 5000
Page 9: ...Revision B MAC 5000 resting ECG analysis system 1 1 2024917 010 1 Introduction ...
Page 10: ...1 2 MAC 5000 resting ECG analysis system Revision B 2024917 010 For your notes ...
Page 21: ...Revision B MAC 5000 resting ECG analysis system 2 1 2024917 010 2 Equipment Overview ...
Page 22: ...2 2 MAC 5000 resting ECG analysis system Revision B 2024917 010 For your notes ...
Page 59: ...Revision B MAC 5000 resting ECG analysis system 3 1 2024917 010 3 Installation ...
Page 60: ...3 2 MAC 5000 resting ECG analysis system Revision B 2024917 010 For your notes ...
Page 77: ...Revision B MAC 5000 resting ECG analysis system 4 1 2024917 010 4 Maintenance ...
Page 78: ...4 2 MAC 5000 resting ECG analysis system Revision B 2024917 010 For your notes ...
Page 103: ...Revision B MAC 5000 resting ECG analysis system 5 1 2024917 010 5 Troubleshooting ...
Page 104: ...5 2 MAC 5000 resting ECG analysis system Revision B 2024917 010 For your notes ...
Page 135: ...Revision B MAC 5000 resting ECG analysis system 6 1 2024917 010 6 Parts List ...
Page 136: ...6 2 MAC 5000 resting ECG analysis system Revision B 2024917 010 For your notes ...
Page 143: ...Revision B MAC 5000 resting ECG analysis system A 1 2024917 010 A Appendix A Abbreviations ...
Page 144: ...A 2 MAC 5000 resting ECG analysis system Revision B 2024917 010 For your notes ...
Page 156: ...B 2 MAC 5000 resting ECG analysis system Revision B 2024917 010 For your notes ...
Page 164: ...C 2 MAC 5000 resting ECG analysis system Revision B 2024917 010 For your notes ...
Page 171: ...Revision B MAC 5000 resting ECG analysis system Index 1 2024917 010 Index ...
Page 172: ...Index 2 MAC 5000 resting ECG analysis system Revision B 2024917 010 ...
Page 174: ...Index Index 4 MAC 5000 resting ECG analysis system Revision B 2024917 010 ...
Page 175: ......