Revision B
3-16
Maintenance: Domestic Electrical Safety Tests
MAC 5000 resting ECG analysis system
2000657-002
Test #3
Patient-cable-leakage-to-ground
Test #4
Patient-cable-leakage-into-patient Leads-from 120 VAC
During this test, line voltage is applied to the patient cable connectors.
To prevent erroneous readings, do not allow the leadwires to contact
conductive materials such as metal handles, and do not place the
leadwires on the floor.
Meter
connectors
V
Gnd
1K
Rvs
Tester
Polarity
Norm
UUT
power
cord
Unit
under
test
(UUT)
Tester
power
cord
Line
Neutral
Gnd
Patient
cable connectors
Patient cable
Line
Neutral
“To be tested” power connector on back of
tester (may not be labeled on some testers).
M13052-03E
Meter
connectors
V
Gnd
1K
Rvs
Tester
Polarity
Norm
UUT
power
cord
Unit
under
test
(UUT)
Tester
power
cord
Line
Neutral
Gnd
Patient
cable connectors
Patient cable
Line
Neutral
“To be tested” power connector on back of
tester (may not be labeled on some testers).
M13052-04E
Summary of Contents for MAC 5000
Page 1: ...MAC 5000 resting ECG analysis system field service manual PN 2000657 002 Revision A...
Page 2: ...T 2...
Page 3: ...field service manual PN 2000657 002 Revision B MAC 5000 resting ECG analysis system...
Page 12: ...Revision B 1 2 MAC 5000 resting ECG analysis system 2000657 002...
Page 20: ...Revision B 2 2 MAC 5000 resting ECG analysis system 2000657 002...
Page 68: ...Revision B 5 2 MAC 5000 resting ECG analysis system 2000657 002...
Page 88: ...Revision B 6 2 MAC 5000 resting ECG analysis system 2000657 002...
Page 114: ...Revision B 7 2 MAC 5000 ECG analysis system 2000657 002...
Page 123: ...Revision B 7 11 PCB Assemblies SD801212 003A CPU MAC 5000 ECG analysis system 2000657 002...
Page 125: ...Revision B 7 13 PCB Assemblies SD801212 003A CPU MAC 5000 ECG analysis system 2000657 002...
Page 127: ...Revision B 7 15 PCB Assemblies SD801212 003A CPU MAC 5000 ECG analysis system 2000657 002...
Page 129: ...Revision B 7 17 PCB Assemblies SD801212 003A CPU MAC 5000 ECG analysis system 2000657 002...
Page 131: ...Revision B 7 19 PCB Assemblies SD801212 003A CPU MAC 5000 ECG analysis system 2000657 002...
Page 133: ...Revision B 7 21 PCB Assemblies SD801212 003A CPU MAC 5000 ECG analysis system 2000657 002...
Page 135: ...Revision B 7 23 PCB Assemblies SD801212 003A CPU MAC 5000 ECG analysis system 2000657 002...
Page 137: ...Revision B 7 25 PCB Assemblies SD801212 003A CPU MAC 5000 ECG analysis system 2000657 002...
Page 142: ...Revision B 7 30 PCB Assemblies SD801222 001A LED MAC 5000 ECG analysis system 2000657 002...
Page 146: ...Revision B A 2 MAC 5000 resting ECG analysis system 2000657 002...
Page 152: ...Revision B A 8 MAC 5000 resting ECG analysis system 2000657 002...
Page 154: ...Revision B B 2 MAC 5000 resting ECG analysis system 2000657 002...