Revision B
3-9
Maintenance: Disassembly Guidelines
MAC 5000 resting ECG analysis system
2000657-002
6. Gently lift the rear of the top cover free from the unit.
7. At the front of the top cover, gently pull the thin strip of plastic
free from under the keyboard. The entire top assembly is now
loose.
8. Carefully lift the top assembly up and clear of the raised display.
Reassembly
1. Raise the display to the vertical position.
2. Make sure the bezel surrounding the rear panel connectors is in
place. Make sure the release mechanism for the Smartmedia
card functions properly.
3. Lower the top cover down around the display and set in
position.
4. Snap the rear of the top cover in place and then, gently pulling
on the thin plastic strip at the front of the top cover, position it
in place under the keyboard assembly.
5. Replace the screws removed in disassembly.
Display/Keyboard
Assembly
Removal
1. Remove the top cover following the procedures above.
2. Label the three cables connecting the display/keyboard
assembly to the main PCB. Disconnect these cables from the
main PCB.
NOTE
The top cover holds the bezel that surrounds
the rear panel connectors, so the bezel may
fall free at this time.
NOTE
It may be helpful to rotate the top cover 45
degrees to provide a larger opening to clear
the display.
NOTE
Two of these cables have locked connectors
that must be lifted up to release the cables.
Summary of Contents for MAC 5000
Page 1: ...MAC 5000 resting ECG analysis system field service manual PN 2000657 002 Revision A...
Page 2: ...T 2...
Page 3: ...field service manual PN 2000657 002 Revision B MAC 5000 resting ECG analysis system...
Page 12: ...Revision B 1 2 MAC 5000 resting ECG analysis system 2000657 002...
Page 20: ...Revision B 2 2 MAC 5000 resting ECG analysis system 2000657 002...
Page 68: ...Revision B 5 2 MAC 5000 resting ECG analysis system 2000657 002...
Page 88: ...Revision B 6 2 MAC 5000 resting ECG analysis system 2000657 002...
Page 114: ...Revision B 7 2 MAC 5000 ECG analysis system 2000657 002...
Page 123: ...Revision B 7 11 PCB Assemblies SD801212 003A CPU MAC 5000 ECG analysis system 2000657 002...
Page 125: ...Revision B 7 13 PCB Assemblies SD801212 003A CPU MAC 5000 ECG analysis system 2000657 002...
Page 127: ...Revision B 7 15 PCB Assemblies SD801212 003A CPU MAC 5000 ECG analysis system 2000657 002...
Page 129: ...Revision B 7 17 PCB Assemblies SD801212 003A CPU MAC 5000 ECG analysis system 2000657 002...
Page 131: ...Revision B 7 19 PCB Assemblies SD801212 003A CPU MAC 5000 ECG analysis system 2000657 002...
Page 133: ...Revision B 7 21 PCB Assemblies SD801212 003A CPU MAC 5000 ECG analysis system 2000657 002...
Page 135: ...Revision B 7 23 PCB Assemblies SD801212 003A CPU MAC 5000 ECG analysis system 2000657 002...
Page 137: ...Revision B 7 25 PCB Assemblies SD801212 003A CPU MAC 5000 ECG analysis system 2000657 002...
Page 142: ...Revision B 7 30 PCB Assemblies SD801222 001A LED MAC 5000 ECG analysis system 2000657 002...
Page 146: ...Revision B A 2 MAC 5000 resting ECG analysis system 2000657 002...
Page 152: ...Revision B A 8 MAC 5000 resting ECG analysis system 2000657 002...
Page 154: ...Revision B B 2 MAC 5000 resting ECG analysis system 2000657 002...