Revision B
2-6
Equipment Overview: Preparation for Use
MAC 5000 resting ECG analysis system
2000657-002
Connector Identification
MD1325-118A
A B C
D
E
F
H
G
I
Table 2-1. Back Panel Connectors
Item
Name
Description
A
A
Not functional.
B
1
Connect a Marquette KISS pump (optional).
C
2
Connect a local transmission cable, serial line,
Marquette KISS pump, or external modem (optional).
D
ANA/TTL
Connect a device requiring analog data or TTL
trigger.
E
EXT.VID.
Connect an external video display.
F
IR
Point at a MAC 5000 or MUSE CV system’s
IR
transceiver to transmit or receive ECG data.
G
card slot
Insert the software card into this slot to run the
system.
H
ground lug
Connect non-grounded peripheral devices to ensure
equipotential.
I
mains AC
power
Insert the mains AC power cable.
WARNING
M15287-7C
Keep leakage current
within acceptable limits
when connecting
auxiliary equipment to
this device.
WARNING
M15287-9D
Total system leakage
current must not
exceed 100
microamperes.
Summary of Contents for MAC 5000
Page 1: ...MAC 5000 resting ECG analysis system field service manual PN 2000657 002 Revision A...
Page 2: ...T 2...
Page 3: ...field service manual PN 2000657 002 Revision B MAC 5000 resting ECG analysis system...
Page 12: ...Revision B 1 2 MAC 5000 resting ECG analysis system 2000657 002...
Page 20: ...Revision B 2 2 MAC 5000 resting ECG analysis system 2000657 002...
Page 68: ...Revision B 5 2 MAC 5000 resting ECG analysis system 2000657 002...
Page 88: ...Revision B 6 2 MAC 5000 resting ECG analysis system 2000657 002...
Page 114: ...Revision B 7 2 MAC 5000 ECG analysis system 2000657 002...
Page 123: ...Revision B 7 11 PCB Assemblies SD801212 003A CPU MAC 5000 ECG analysis system 2000657 002...
Page 125: ...Revision B 7 13 PCB Assemblies SD801212 003A CPU MAC 5000 ECG analysis system 2000657 002...
Page 127: ...Revision B 7 15 PCB Assemblies SD801212 003A CPU MAC 5000 ECG analysis system 2000657 002...
Page 129: ...Revision B 7 17 PCB Assemblies SD801212 003A CPU MAC 5000 ECG analysis system 2000657 002...
Page 131: ...Revision B 7 19 PCB Assemblies SD801212 003A CPU MAC 5000 ECG analysis system 2000657 002...
Page 133: ...Revision B 7 21 PCB Assemblies SD801212 003A CPU MAC 5000 ECG analysis system 2000657 002...
Page 135: ...Revision B 7 23 PCB Assemblies SD801212 003A CPU MAC 5000 ECG analysis system 2000657 002...
Page 137: ...Revision B 7 25 PCB Assemblies SD801212 003A CPU MAC 5000 ECG analysis system 2000657 002...
Page 142: ...Revision B 7 30 PCB Assemblies SD801222 001A LED MAC 5000 ECG analysis system 2000657 002...
Page 146: ...Revision B A 2 MAC 5000 resting ECG analysis system 2000657 002...
Page 152: ...Revision B A 8 MAC 5000 resting ECG analysis system 2000657 002...
Page 154: ...Revision B B 2 MAC 5000 resting ECG analysis system 2000657 002...