9-24
MAC™ 1600
2028451-182B
System Configuration
Patient Setup
The
Patient Setup
function allows you to define the:
Available and required patient information
Available test information
Available clinical trial information
Only if the
CTDG CT Data Guard
option is activated.
Barcode reader settings
Only if the
BCRD
USB Barcode Reader
option is activated
To access
Patient Setup
from the MAC 1600
Main Menu
, press
F5
(
System Configuration
) >
F6
(
More
) >
F4
(
Patient Setup
).
The following table identifies the settings on
Patient Setup
.
ST Level Unit
Determines whether the ST segment will be measured in
millivolts (mV) or millimeters (mm).
Blood Pressure Unit
Determines whether blood pressure will be measured in
millimeters of mercury (mmHg) or kilopascals (kPa).
Line Filter
Determines the frequency of the line filter. Options are
50 Hz and 60 Hz.
Lead Label
Determines whether the system will label leads using the
standards of the International Electrotechnical Commission
(IEC) or the American Heart Association (AHA).
Field Comments
Field
Comment
Patient Information Setup Window
Patient ID
Determines whether the patient ID is required. On reports,
it will be labelled ID.
Secondary ID
Determines whether a secondary patient ID will be
available when entering patient data and whether it is
required. It can only be required if it is first enabled. On
reports, it will be labelled ID 2.
Last Name
Determines whether the patient’s last name will be
available when entering patient data and whether it is
required. It can only be required if it is first enabled.
First Name
Determines whether the patient’s first name will be
available when entering patient data and whether it is
required. It can only be required if it is first enabled.
Kanji Name
Determines whether the Kanji name will be available when
entering patient data.
Summary of Contents for MAC 1600
Page 8: ...vi MAC 1600 2028451 182B...
Page 9: ...2028451 182B MAC 1600 1 1 1 Introduction...
Page 24: ...1 16 MAC 1600 2028451 182B Introduction...
Page 25: ...2028451 182B MAC 1600 2 1 2 Equipment Overview...
Page 38: ...2 14 MAC 1600 2028451 182B Equipment Overview...
Page 39: ...2028451 182B MAC 1600 3 1 3 Preparing the Patient...
Page 46: ...3 8 MAC 1600 2028451 182B Preparing the Patient...
Page 47: ...2028451 182B MAC 1600 4 1 4 Entering Patient Information...
Page 50: ...4 4 MAC 1600 2028451 182B Entering Patient Information...
Page 51: ...2028451 182B MAC 1600 5 1 5 Recording a Resting ECG...
Page 65: ...2028451 182B MAC 1600 6 1 6 Arrhythmia Mode Recording...
Page 71: ...2028451 182B MAC 1600 7 1 7 Stress Testing...
Page 82: ...7 12 MAC 1600 2028451 182B Stress Testing...
Page 83: ...2028451 182B MAC 1600 8 1 8 Managing Internal Storage...
Page 93: ...2028451 182B MAC 1600 9 1 9 System Configuration...
Page 128: ...9 36 MAC 1600 2028451 182B System Configuration...
Page 129: ...2028451 182B MAC 1600 10 1 10 Maintenance...
Page 143: ...2028451 182B MAC 1600 A 1 A Troubleshooting...
Page 151: ...2028451 182B MAC 1600 B 1 B Creating Barcodes...
Page 161: ......