System Configuration
2028451-182B
MAC™ 1600
9-11
1–12 Lead
Twelve fields that allow you to define the sequence in
which the leads will appear. Available only if SEQ4 is
selected for the Lead Sequence.
1–12 Label
Twelve fields that allow you to define the labels that will
appear/print for the corresponding leads. Available only if
SEQ4 is selected for the Lead Sequence.
1–6 Rhythm Leads
Six fields that allow you to define the rhythm leads and
their sequence. You can select the rhythm leads for all four
lead sequences.
Page 4
10s ECG Report Format
Determines how the 10s ECG report will print. If no format
is selected, the report will not print.
Detailed Results Report
Format
Determines how the Detailed Results report will print. If no
format is selected, the report will not print.
Report Copies
Determines how many copies of the selected report will
print.
Print Interpretation
Determines whether ECG interpretation will print on the
report. Available only if either the MI12 or MIHR option is
activated. For more information, refer to
Auto Store ECG
Determines whether the ECG will automatically be stored
on the internal storage. Available only if the M100 internal
storage option is activated. For more information, refer to
File Manager Sort By
Determines the field by which the File Manager will sort
records in internal storage. Available only if the M100
internal storage option is activated.
Auto Transmit ECG
Determines whether the ECG will automatically be
transmitted to an external device. Available only if one of
the communications options is activated.
For more information, refer to
Field
Comment
Summary of Contents for MAC 1600
Page 8: ...vi MAC 1600 2028451 182B...
Page 9: ...2028451 182B MAC 1600 1 1 1 Introduction...
Page 24: ...1 16 MAC 1600 2028451 182B Introduction...
Page 25: ...2028451 182B MAC 1600 2 1 2 Equipment Overview...
Page 38: ...2 14 MAC 1600 2028451 182B Equipment Overview...
Page 39: ...2028451 182B MAC 1600 3 1 3 Preparing the Patient...
Page 46: ...3 8 MAC 1600 2028451 182B Preparing the Patient...
Page 47: ...2028451 182B MAC 1600 4 1 4 Entering Patient Information...
Page 50: ...4 4 MAC 1600 2028451 182B Entering Patient Information...
Page 51: ...2028451 182B MAC 1600 5 1 5 Recording a Resting ECG...
Page 65: ...2028451 182B MAC 1600 6 1 6 Arrhythmia Mode Recording...
Page 71: ...2028451 182B MAC 1600 7 1 7 Stress Testing...
Page 82: ...7 12 MAC 1600 2028451 182B Stress Testing...
Page 83: ...2028451 182B MAC 1600 8 1 8 Managing Internal Storage...
Page 93: ...2028451 182B MAC 1600 9 1 9 System Configuration...
Page 128: ...9 36 MAC 1600 2028451 182B System Configuration...
Page 129: ...2028451 182B MAC 1600 10 1 10 Maintenance...
Page 143: ...2028451 182B MAC 1600 A 1 A Troubleshooting...
Page 151: ...2028451 182B MAC 1600 B 1 B Creating Barcodes...
Page 161: ......