Measurement and Analysis
LOGIQ V2/LOGIQ V1
–
User Guide
2-55
Direction 5610736-100
Rev. 9
Intensity (Echo level) measurement
To make an echo level measurement:
1. Press
Measure
key to enter into measure mode, if
necessary.
2. Adjust
Caliper Change
in the Primary Menu to
Intensity
.
NOTE:
The available values for
Caliper Change
menu can be set
in Utility -> Measure -> Advanced preset.
3. To position the caliper at the start point, move the
Trackball
.
4. To fix the trace start point, press
Set
. The caliper changes to
an active caliper.
5. To trace the measurement area, move the
Trackball
around
the anatomy. A dotted line shows the traced area.
6. To complete the measurement, press
Set
. The system
displays the echo level, as EL __ dB, in the Results Window.
NOTE:
The echo level measurement is only available on a frozen
image, not on a B-paused image.