LOGIQ Totus
Basic Service Manual
Chapter 3 LOGIQ Totus Setup
3 - 17
3-5-6 Connecting probes (continued)
To disconnect a probe:
1.) Select a different probe or freeze the image before removing a probe to avoid disconnecting a live
probe.
2.) Rotate the lock handle counter-clockwise to the horizontal position to unlock the connector.
3.) Remove the connector from the port.
4.) Ensure that the probe head is clean before placing the probe in its storage case. For cleaning
instructions, “Probe Cleaning Process” in Chapter 11 of the LOGIQ Totus Basic User
Manual (latest revision).