Section 6-4 - DICOM Server Service Adjustments
Section 6-4
DICOM Server Service Adjustments
6-4-1
Slow response rate between the LOGIQ Totus and a DICOM (PACS) server
Problems with slow responses may result in images being re-sent automatically and low transfer rates.
If you are experiencing problems with slow responses from a DICOM server, increase the timeout
setting.
To adjust properties for a DICOM server (online)
:
1.) Navigate to
Utility
>
Connectivity
>
Dicom > Service
.
2.) Select the service for the DICOM server.
3.) Under
Connection
, set
Timeout (sec)
to
360
.
6-4-2
Shutdown problems when running a portable LOGIQ Totus
When portable (offline and not communicating with a PACS server), use the following settings to avoid
problems with shutdown speed:
To adjust properties for a DICOM server (offline)
:
1.) Navigate to
Utility
>
Connectivity
>
Dicom > Service
.
2.) Select the service for the DICOM server.
3.) Under
Connection
, set these parameters:
-
Maximum Retries
- Set to
0
.
-
Retry Interval (sec)
- Set to the lowest possible value.
-
Timeout (sec)
- Set to the lowest possible value.
Section 6-5
Volume Navigation (VNAV) Calibration
There is no calibration available in the field. If the registration ms to be out of calibration, reloading the
software will restore the factory settings. If the registration still ms out of calibration, replace the Volume
Navigation (VNAV) hardware.
For information on the hardware replacement,
8-11-19 "Replacement of the VNAV Option Cable Set-U"
.
Summary of Contents for LOGIQ Totus
Page 2: ......
Page 4: ...GE HealthCare CONFIDENTIAL Direction 5936427 Rev 1 LOGIQ Totus Basic Service Manual ii...
Page 5: ...GE HealthCare CONFIDENTIAL Direction 5936427 Rev 1 LOGIQ Totus Basic Service Manual iii...
Page 6: ...GE HealthCare CONFIDENTIAL Direction 5936427 Rev 1 LOGIQ Totus Basic Service Manual iv...
Page 643: ......