Section 4-3 - Functional Checks
4-3-8 B-Mode functional checks (continued)
Probe indications for use
the online Help or Chapter 11 of the appropriate version of the LOGIQ Totus Basic User Manual.
1 "Label Location" on page 1-13
Preparations
You may use a phantom (optional) when doing these tests.
1.) Connect one of the probes, to the scanner’s active probe connector.
-
Section 3-5-6 "Connecting probes" on page 3-16
for info about connecting the probes.
-
For available probes,
Section 9-13 "Probes" on page 9-24
2.) Turn ON the scanner. The B-Mode window is displayed (default mode).
3.) If needed, adjust the Brightness and Contrast setting on the monitor (
).
4.) Press
B-MODE
on the Operator Panel to access B-Mode.
5.) These Image Controls are used to optimize the B-Mode picture:
-
Use
Gain
and
TGC
controls to optimize the overall image together with the
Power
control.
-
Use
Depth
to adjust the range to be imaged.
-
Use
Frequency
(move to higher frequencies) or
Frame rate
(move to lower frame rate) to
increase resolution in image.
-
Use
Frequency
(move to lower frequency) to increase penetration.
-
Use the control to optimize imaging in the blood flow regions and make a cleaner, less noisy
image.
-
Use
Reject
controls to reduce noise in the image.
Checks
•
Check Width, Frame rate, Frequency. The results of these adjustments must be verified on the
B-Mode sector on the screen.
•
Check Up/Down, Left/Right, B Color Maps and Cineloop. The results of these adjustments must be
verified on the B-Mode sector on the screen.
•
Check Gain, TGC and Depth.
•
Check B-Mode Soft Menu Controls.
•
Check Compress, Contour, Reject and Tilt.
•
Check Power and Dynamic Range.
WARNING
WARNING
ALWAYS USE THE MINIMUM POWER REQUIRED TO OBTAIN ACCEPTABLE IMAGES
IN ACCORDANCE WITH APPLICABLE GUIDELINES AND POLICIES.
Summary of Contents for LOGIQ Totus
Page 2: ......
Page 4: ...GE HealthCare CONFIDENTIAL Direction 5936427 Rev 1 LOGIQ Totus Basic Service Manual ii...
Page 5: ...GE HealthCare CONFIDENTIAL Direction 5936427 Rev 1 LOGIQ Totus Basic Service Manual iii...
Page 6: ...GE HealthCare CONFIDENTIAL Direction 5936427 Rev 1 LOGIQ Totus Basic Service Manual iv...
Page 643: ......