Section 3-5 - Completing the Setup
3-5-16
Complete power down
1.) Press once on the
ON
/
OFF
button on the operator panel to display the
SYSTEM - EXIT
menu
2.) Select
Shutdown
to do a complete power down of the LOGIQ Totus.
The back end will first turn off the LOGIQ Totus activity and print the message “Please wait -
Shutdown in progress” in the display on the operator panel.
Next, it starts to shut down. The time to turn down the LOGIQ Totus, including the back end, may
vary from 10 seconds up to approximately 1 minute.
The last thing that shuts down, is the light on the operator panel, indicating that you can continue
with the next step.
3.) Switch off the Mains Power Circuit Breaker, located on the rear of the LOGIQ Totus. This will cut
power distribution within the LOGIQ Totus.
WARNING
WARNING
RISK OF LOSING DATA. BE SURE TO WAIT UNTIL THE LOGIQ TOTUS HAS FINISHED
ITS SHUT-DOWN. FAILING TO DO SO MAY CORRUPT DATA ON THE HDD, MAKING
THE LOGIQ TOTUS FAIL.