GE H
EALTHCARE
PROPRIETARY TO GE
D
IRECTION
5394227, 12
LOGIQ S8/LOGIQ E8 S
ERVICE
M
ANUAL
Section 3-6 - Completing the Set Up
3 - 15
3-6-4
Connecting Probes
3-6-4-1
Introduction
Probes can be connected at any time, regardless of whether the unit is on or off. To ensure that the
ports are not active, place the system in the image freeze condition
.
Figure 3-7 Probe port
3-6-4-2
Cable Handling
Take the following precautions with probe cables:
•
Keep free from wheels
•
Do not bend the cable acutely
•
Avoid crossing cables between probes.
3-6-4-3
Connect a probe
1.)
Before connecting the probe:
a.) Do a visual check of the probe pins and system sockets.
b.) Remove any dust or foam rests from the probe pins.
c.) Inspect the probe for damage or degradation to the housing, strain relief, lens, seal, cable and
connector.
2.)
Put the probe in the probe holder. Use the integrated cable management hook to wrap the cord.
3.)
Ensure that the probe locking handle is at horizontal position.
4.)
Hold the probe connector vertically with the cable pointing upward.
5.)
Align the connector with the probe port and carefully push into place.
6.)
Turn the connector locking handle clockwise (to vertical position) to secure the probe connector.
7.)
Carefully position the probe cable so it is free to move and is not resting on the floor.
CAUTION
!! CAUTION:
DO NOT use a probe which appears damaged until functional and safe performance is
verified.
CAUTION
!! CAUTION:
Fault conditions can result in electric shock hazard. Do not touch the surface of probe
connectors which are exposed when the probe is removed.
CAUTION
!! CAUTION:
DO NOT allow the probe head to hang free. Impact to the probe head could result in
irreparable damage.
a. Active probe port
b. Parking probe port
c. Pencil probe port
LOGIQ S8/LOGIQ S8 Vet
LOGIQ E8
a
b
Summary of Contents for LOGIQ S8
Page 11: ...GE HEALTHCARE DIRECTION 5394227 12 LOGIQ S8 LOGIQ E8 SERVICE MANUAL ix ZH CN KO...
Page 12: ...GE HEALTHCARE DIRECTION 5394227 12 LOGIQ S8 LOGIQ E8 SERVICE MANUAL x...
Page 32: ...GE HEALTHCARE DIRECTION 5394227 12 LOGIQ S8 LOGIQ E8 SERVICE MANUAL 16...
Page 416: ...GE HEALTHCARE DIRECTION 5394227 REVISION 12 LOGIQ S8 LOGIQ E8 SERVICE MANUAL Index 4...
Page 417: ......