GE
RAFT
LOGIQ P9/P7
D
IRECTION
5604324
, R
EVISION
11
DRAFT (J
ANUARY
24, 2019)
S
ERVICE
M
ANUAL
8-164
Section 8-37 - Replacement of the Monitor Arm
Section 8-37
Replacement of the Monitor Arm
8-37-1
Manpower
1 person, 15 minutes.
8-37-2
Tools
Standard Phillips Screwdriver, Nipper, 4 mm Hex Wrench, 3 Cable ties (GEPN: 2160955).
8-37-3
Removal Procedure
1) Remove LCD monitor and LCD Arm Plastic Covers.(refer to
)
2) Remove Harness Cable Assy and Monitor Arm. (refer to
1)~4))
8-37-4
Installation Procedure
In reverse order of removal.
8-37-5
Functional Check
1) Visually check LCD monitor
2) LCD Arm Test. (Refer to