Section 8-28 - Replacement of the Harness Cable and OPIO Cable Assy
Figure 8-167 Removal of Harness Cable Assy
8-28-5
Removal Procedure (OPIO Cable Assy)
1) Unscrew 1 screw, remove cable clamp, and then loosen cable through hole.
NOTE:
There is cable position marker at this location.
2) Release the cable holder and carefully pull down cables.
3) Unscrew 2 screws and remove Link arm cable fix brkt.
4) Carefully pull up cables and remove cables
Figure 8-168 Removal of Harness Cable Assy
8-28-6
Installation Procedure
Install components in reverse order of removal. Replace the cut cable ties with new ones of the