GE
DRAFT
LOGIQ P9/P7
D
IRECTION
5604324
, R
EVISION
11
DRAFT (J
ANUARY
24, 2019)
S
ERVICE
M
ANUAL
Chapter 8 - Replacement Procedures
8-7
Section 8-3
System Software - Installation/Upgrade Procedure
8-3-1
Introduction
This section describes procedure to install/upgrade software to system.
8-3-2
Manpower
1 person, 1 hour
- Depends on contents of System DVD, peripherals, etc. (R1.x.x or R2.x.x)
- Depends on contents of USB media, peripherals, etc. (R3.x.x or later)
8-3-3
Tools
- System DVD (R1.x.x or R2.x.x)
- Software Flash Drive with Base Load Image and Application Software. (R3.x.x or later)
8-3-4
Preparations
Before performing the Software Upgrade:
•
make sure that all system functions are working correct
•
check the current Application Software version and the installed Options
8-3-5
Preparation
8-3-5-1
General Notes on LOGIS P7/P9 Software Loading
When preparing to load software, you must perform the following steps:
•
Back up the entire Patient Archive and Reports (you can import the database from previous version
into this version).
•
Back up all presets, including Service.
•
Record Connectivity settings (TCP/IP).
•
Load Base System Software and install Application Software:
-
Reload entire drive (all partitions) or
-
Reload only C partition