Section 7-4 - Screen Captures and Logs
7-4-3-3
System voltage failure
7-4-3-3-1
Cause
System detected unexpected voltage level from voltage level monitoring circuits.
7-4-3-3-2
What to Do
1) Press OK and reboot the system.
2) If problem persists, then shut down the system, turn OFF breaker and reboot.
7-4-3-3-3
Log message of the Voltage error
(Example)
Error ; acqHouseKeeping(540); MCB_AFE_3V3_0= 2.5 outside tolerance limits. UpperBound =
3.6 LowerBound = 2.970, Soft Limits are usually narrower than upperBound and lower bound