Performing an Exam
2-30
LOGIQ E9
–
User Guide
Direction 5454884-100 English
Rev. 1
Time interval
To measure a horizontal time interval and velocity:
1. Press
Measure
. Press the top Trackball key to select Time;
an active caliper with vertical and horizontal dotted lines
displays.
2. To position the caliper at the start point, move the
Trackball
.
3. To fix the first caliper, press
Set
. The system fixes the first
caliper and displays a second active caliper.
4. To position the second caliper at the end point, move the
Trackball
.
5. To complete the measurement, press
Set
. The system
displays the time interval between the two calipers in the
Results Window.
Slope (Time interval and Velocity)
To measure time and velocity between two points:
1. Press
Measure
. Press the top Trackball key to select Slope;
an active caliper with vertical and horizontal dotted lines
displays.
2. To position the active caliper at the start point, move the
Trackball
.
3. To fix the start point, press
Set
.
The system fixes the first caliper and displays a second
active caliper.
4. To position the second caliper at the end point, move the
Trackball
.
5. To complete the measurement, press
Set
.
The system displays time(s) and slope between the two
points in the Results Window.
Summary of Contents for LOGIQ E9
Page 8: ...i 6 LOGIQ E9 User Guide Direction 5454884 100 English Rev 1 ...
Page 12: ...i 10 LOGIQ E9 User Guide Direction 5454884 100 English Rev 1 ...
Page 126: ...Performing an Exam 2 40 LOGIQ E9 User Guide Direction 5454884 100 English Rev 1 ...
Page 202: ...After the Exam is Over 3 76 LOGIQ E9 User Guide Direction 5454884 100 English Rev 1 ...
Page 250: ...Safety 4 48 LOGIQ E9 User Guide Direction 5454884 100 English Rev 1 ...
Page 254: ...Index 4 LOGIQ E9 User Guide Direction 5454884 100 English Rev 1 ...
Page 255: ......
Page 256: ......