System maintenance
LOGIQ e – Basic Service Manual
10-19
5461614-100 English Rev. 6
Functional checks
NOTE:
See also Chapter 4
The functional checks take about 60 minutes to perform. Refer
to the Ultrasound system user documentation whenever
necessary.
System checks
Table 10-8: System functional checks
Step
Item
Description
1.
B-Mode
Verify basic B-Mode (2D) operation. Check the basic Ultrasound system
controls that affect this mode of operation.
2.
CF-Mode
Verify basic CF-Mode (Color Flow Mode) operation. Check the basic
Ultrasound system controls that affect this mode of operation.
3.
Doppler Modes
Verify basic Doppler operation (PW and CW if available). Check the basic
Ultrasound system controls that affect this mode of operation.
4.
M-Mode
Verify basic M-Mode operation. Check the basic Ultrasound system
controls that affect this mode of operation.
5.
Probe Elements
Perform an Element Test on each probe to verify that all the probe
elements and system channels are functional.
6. Applicable
Software Options
Verify the basic operation of all optional modes such as Contrast. Check
the basic Ultrasound system controls that affect each options operation.
7. Xmit/Recv
Elements
Use the Visual Channel Utility on the loop connect to verify that all system
xmit/recv channels are functional.
8. Operator
Panel
test
Perform the Operator Panel Test Procedure.
9.
Keyboard
Do the interactive keyboard test.
10.
LCD
Verify basic LCD display functions. Refer to Chapter 3 of the User Manual.
11. Software
Menu
check
Verify Software Menu display functions. Refer to Chapter 3 of the User
Manual.
12. Peripherals
See:
13.
Measurements
In measurement mode, make distance measurement, get result in result
window. Verify the distance by graduate rule. Distance Accuracy should
be within
±
5%. (Name result from result window Result A, result from
graduate rule Result B; Distance Accuracy = (Result B-Result A)/Result A)
Summary of Contents for LOGIQ e
Page 5: ...LOGIQ e Basic Service Manual i 3 5461614 100 English Rev 6...
Page 6: ...i 4 LOGIQ e Basic Service Manual 5461614 100 English Rev 6...
Page 7: ...LOGIQ e Basic Service Manual i 5 5461614 100 English Rev 6...
Page 8: ...i 6 LOGIQ e Basic Service Manual 5461614 100 English Rev 6...
Page 9: ...LOGIQ e Basic Service Manual i 7 5461614 100 English Rev 6...
Page 10: ...i 8 LOGIQ e Basic Service Manual 5461614 100 English Rev 6...
Page 11: ...LOGIQ e Basic Service Manual i 9 5461614 100 English Rev 6...
Page 12: ...i 10 LOGIQ e Basic Service Manual 5461614 100 English Rev 6...
Page 24: ...i 22 LOGIQ e Basic Service Manual 5461614 100 English Rev 6...
Page 60: ...Introduction 1 36 LOGIQ e Basic Service Manual 5461614 100 English Rev 6...
Page 80: ...Site Preparations 2 20 LOGIQ e Basic Service Manual 5461614 100 English Rev 6...
Page 134: ...System Setup 3 54 LOGIQ e Basic Service Manual 5461614 100 English Rev 6...
Page 208: ...Components and Functions Theory 5 14 LOGIQ e Basic Service Manual 5461614 100 English Rev 6...
Page 262: ...Replacement Procedures 8 26 LOGIQ e Basic Service Manual 5461614 100 English Rev 6...
Page 328: ...Care and Maintenance 10 44 LOGIQ e Basic Service Manual 5461614 100 English Rev 6...
Page 384: ...Docking Cart Setup 11 56 LOGIQ e Basic Service Manual 5461614 100 English Rev 6...
Page 426: ...Index 4 LOGIQ e Basic Service Manual 5461614 100 English Rev 6...
Page 427: ...GE...