OB Measurements
170
LOGIQ
t
α
100 User Manual
2211157–100 Rev 0
Velocity (mm/second)
VELOCITY is the difference in tissue depth (range of
tissue motion) divided by the elapsed time captured in
the M-Mode image or B/M-Mode. If calculations are
done in the wrong mode, an error message VEL ILG will
appear in the Calculation Result Area. Velocity
calculations are measured as moving mm/second.
Velocity= Height of a Wave/Elapsed Time.
Select [4] in the Sub Menu. If the sub menu was used for
a prior calculation, the system defaults back to the sub
menu when
MEASUREMENT is pressed.
Use the
TRACKBALL to move the M-Mode measurement
cursor to the start point of the measurement.
Press
SET to fix the start point. A second M-Mode
measurement cursor appears.
Use the
TRACKBALL to move the cursor to the end point
of the measurement.
Press
SET to fix the end point. Should further
measurements need to be taken, press
SET again to
invoke the second cursor and follow the procedure as
indicated above. Four calculations can be taken at the
same time.
+
+
4
TB
TB
Summary of Contents for LOGIQ A100
Page 18: ...Introduction LOGIQt α100 User Manual 2211157 100 Rev 0 2 This page left blank intentionally ...
Page 54: ...Safety LOGIQt α100 User Manual 2211157 100 Rev 0 38 This page left blank intentionally ...
Page 74: ...Device Labels 58 LOGIQt α100 User Manual 2211157 100 Rev 0 This page left blank intentionally ...
Page 76: ...Safety LOGIQt α100 User Manual 2211157 100 Rev 0 60 This page left blank intentionally ...
Page 88: ...Image Display 72 LOGIQt α100 User Manual 2211157 100 Rev 0 This page left blank intentionally ...
Page 146: ...OB LOGIQt α100 User Manual 2211157 100 Rev 0 130 This page left blank intentionally ...
Page 236: ...Cardiology 220 LOGIQt α100 User Manual 2211157 100 Rev 0 This page left blank intentionally ...
Page 248: ...Urology LOGIQt α100 User Manual 2211157 100 Rev 0 232 This page left blank intentionally ...
Page 251: ...LOGIQt α100 User Manual 2211157 100 Rev 0 235 Control Keys Configuration Using Control Keys ...
Page 252: ...Control Keys LOGIQt α100 User Manual 2211157 100 Rev 0 236 This page left blank intentionally ...
Page 298: ...Probes 282 LOGIQt α100 User Manual 2211157 100 Rev 0 This page left blank intentionally ...
Page 344: ...Maintenance 328 LOGIQt α100 User Manual 2211157 100 Rev 0 This page left blank intentionally ...
Page 346: ...OB Tables LOGIQt α100 User Manual 2211157 100 Rev 0 330 This page left blank intentionally ...
Page 400: ...OB Tables 384 LOGIQt α100 User Manual 2211157 100 Rev 0 This page left blank intentionally ...
Page 408: ...Index Index 8 LOGIQt α100 User Manual 2211157 100 Rev 0 This page left blank intentionally ...