Basic Measurements
119
LOGIQ
t
α
100 User Manual
2211157–100 Rev 0
Velocity Measurement
Velocity of a moving structure is the rate of change of its
displacement. It is determined by placing an arbitrary
line tangent to the steepest slope of the structure in a
M–Mode trace and measuring its slope which is equal to
the velocity of the structure. It is expressed in meters per
second.
As in all M–Mode measurement freeze before performing
any measurement.
Press
MEASUREMENT to display the General
Measurement Menu. Simultaneously an open ended “ ”
appears on the screen.
Select [3].
If the appropriate measurement was already performed
the Velocity is calculated and displayed.
Use the
TRACKBALL to move the open ended “ ” cursor
to a point tangential to the steepest slope of the M–Mode
trace.
Press
SET set to fix one end of the line. A second cursor
and a dotted line appears.
Use the
TRACKBALL to move the second open ended
“ ” cursor to a point such that the dotted line is tangential
to the steepest slope of the M–Mode trace of the
structure of interest.
Press
SET. The velocity of the structure is displayed at
the right side off the screen.
Up to four measurements values can be displayed on
the screen.
+
+
A
"
TB
TB
3
Meas
Summary of Contents for LOGIQ A100
Page 18: ...Introduction LOGIQt α100 User Manual 2211157 100 Rev 0 2 This page left blank intentionally ...
Page 54: ...Safety LOGIQt α100 User Manual 2211157 100 Rev 0 38 This page left blank intentionally ...
Page 74: ...Device Labels 58 LOGIQt α100 User Manual 2211157 100 Rev 0 This page left blank intentionally ...
Page 76: ...Safety LOGIQt α100 User Manual 2211157 100 Rev 0 60 This page left blank intentionally ...
Page 88: ...Image Display 72 LOGIQt α100 User Manual 2211157 100 Rev 0 This page left blank intentionally ...
Page 146: ...OB LOGIQt α100 User Manual 2211157 100 Rev 0 130 This page left blank intentionally ...
Page 236: ...Cardiology 220 LOGIQt α100 User Manual 2211157 100 Rev 0 This page left blank intentionally ...
Page 248: ...Urology LOGIQt α100 User Manual 2211157 100 Rev 0 232 This page left blank intentionally ...
Page 251: ...LOGIQt α100 User Manual 2211157 100 Rev 0 235 Control Keys Configuration Using Control Keys ...
Page 252: ...Control Keys LOGIQt α100 User Manual 2211157 100 Rev 0 236 This page left blank intentionally ...
Page 298: ...Probes 282 LOGIQt α100 User Manual 2211157 100 Rev 0 This page left blank intentionally ...
Page 344: ...Maintenance 328 LOGIQt α100 User Manual 2211157 100 Rev 0 This page left blank intentionally ...
Page 346: ...OB Tables LOGIQt α100 User Manual 2211157 100 Rev 0 330 This page left blank intentionally ...
Page 400: ...OB Tables 384 LOGIQt α100 User Manual 2211157 100 Rev 0 This page left blank intentionally ...
Page 408: ...Index Index 8 LOGIQt α100 User Manual 2211157 100 Rev 0 This page left blank intentionally ...