Basic Measurements
116
LOGIQ
t
α
100 User Manual
2211157–100 Rev 0
Measurement of Volume by approximation to a Spheroid (Three
Distance Method)
This formula is suited for anatomical structures that are
arbitrarily shaped and do not conform to radial or bilateral
symmetries.
Pre Selection
Using the
LEFT/RIGHT keys scan and obtain a section
of the organ in a B/B image. For details of using the B/B
image please refer page 71.
Measure its long and short axes by distance method.
OR
Construct an ellipse to approximate its outline.
Turn the probe through 90
°
and obtain another section.
Press the
RIGHT/LEFT keys to freeze in the second B/B
image. For details refer page 71.
Measure the third axis.
OR
Construct an ellipse to approximate the outline.
NOTE: Of the two axes of the second ellipse displayed
on measurement area on the side, ensure that the third
axis is always on the left. To swap the axes press
measurement key.
Select
2. The volume is calculated and displayed in the
calculation result area. Distances d
1
,d
2
,d
3
used in
calculation are also displayed.
2
OR
OR
ÀÀ
ÀÀ
L
R
ÀÀ
ÀÀ
R
L
Summary of Contents for LOGIQ A100
Page 18: ...Introduction LOGIQt α100 User Manual 2211157 100 Rev 0 2 This page left blank intentionally ...
Page 54: ...Safety LOGIQt α100 User Manual 2211157 100 Rev 0 38 This page left blank intentionally ...
Page 74: ...Device Labels 58 LOGIQt α100 User Manual 2211157 100 Rev 0 This page left blank intentionally ...
Page 76: ...Safety LOGIQt α100 User Manual 2211157 100 Rev 0 60 This page left blank intentionally ...
Page 88: ...Image Display 72 LOGIQt α100 User Manual 2211157 100 Rev 0 This page left blank intentionally ...
Page 146: ...OB LOGIQt α100 User Manual 2211157 100 Rev 0 130 This page left blank intentionally ...
Page 236: ...Cardiology 220 LOGIQt α100 User Manual 2211157 100 Rev 0 This page left blank intentionally ...
Page 248: ...Urology LOGIQt α100 User Manual 2211157 100 Rev 0 232 This page left blank intentionally ...
Page 251: ...LOGIQt α100 User Manual 2211157 100 Rev 0 235 Control Keys Configuration Using Control Keys ...
Page 252: ...Control Keys LOGIQt α100 User Manual 2211157 100 Rev 0 236 This page left blank intentionally ...
Page 298: ...Probes 282 LOGIQt α100 User Manual 2211157 100 Rev 0 This page left blank intentionally ...
Page 344: ...Maintenance 328 LOGIQt α100 User Manual 2211157 100 Rev 0 This page left blank intentionally ...
Page 346: ...OB Tables LOGIQt α100 User Manual 2211157 100 Rev 0 330 This page left blank intentionally ...
Page 400: ...OB Tables 384 LOGIQt α100 User Manual 2211157 100 Rev 0 This page left blank intentionally ...
Page 408: ...Index Index 8 LOGIQt α100 User Manual 2211157 100 Rev 0 This page left blank intentionally ...