Basic Measurements
100
LOGIQ
t
α
100 User Manual
2211157–100 Rev 0
Measurement Accuracy
The following information is intended to provide guidance
to the user in determining the amount of variation or
measurement error that should be considered when
performing clinical measurements with this equipment.
Error can be contributed by equipment limitations and
improper user technique. Be sure to follow all
measurement instructions and develop uniform
measurement techniques among all users to minimize
the potential operator error. Also, in order to detect
possible equipment malfunctions that could affect
measurement accuracy, a quality assurance plan should
be established for the equipment that includes routine
accuracy checks.
Measurement
Accuracy
Range
Depth
"
1mm
Full screen
Distance
"
1mm
Full screen
Circumference
D
Trace
D
Ellipse
"
1mm
Full screen
Area
D
Trace
D
Ellipse
"
1mm
2
Full screen
Time (seconds)
"
5%
Time Line Display
Table 8. Measurement Accuracy
Distance and Circumference measurements are
displayed in mm and Area measurements in cm
2
.
.
Summary of Contents for LOGIQ A100
Page 18: ...Introduction LOGIQt α100 User Manual 2211157 100 Rev 0 2 This page left blank intentionally ...
Page 54: ...Safety LOGIQt α100 User Manual 2211157 100 Rev 0 38 This page left blank intentionally ...
Page 74: ...Device Labels 58 LOGIQt α100 User Manual 2211157 100 Rev 0 This page left blank intentionally ...
Page 76: ...Safety LOGIQt α100 User Manual 2211157 100 Rev 0 60 This page left blank intentionally ...
Page 88: ...Image Display 72 LOGIQt α100 User Manual 2211157 100 Rev 0 This page left blank intentionally ...
Page 146: ...OB LOGIQt α100 User Manual 2211157 100 Rev 0 130 This page left blank intentionally ...
Page 236: ...Cardiology 220 LOGIQt α100 User Manual 2211157 100 Rev 0 This page left blank intentionally ...
Page 248: ...Urology LOGIQt α100 User Manual 2211157 100 Rev 0 232 This page left blank intentionally ...
Page 251: ...LOGIQt α100 User Manual 2211157 100 Rev 0 235 Control Keys Configuration Using Control Keys ...
Page 252: ...Control Keys LOGIQt α100 User Manual 2211157 100 Rev 0 236 This page left blank intentionally ...
Page 298: ...Probes 282 LOGIQt α100 User Manual 2211157 100 Rev 0 This page left blank intentionally ...
Page 344: ...Maintenance 328 LOGIQt α100 User Manual 2211157 100 Rev 0 This page left blank intentionally ...
Page 346: ...OB Tables LOGIQt α100 User Manual 2211157 100 Rev 0 330 This page left blank intentionally ...
Page 400: ...OB Tables 384 LOGIQt α100 User Manual 2211157 100 Rev 0 This page left blank intentionally ...
Page 408: ...Index Index 8 LOGIQt α100 User Manual 2211157 100 Rev 0 This page left blank intentionally ...