GE H
EALTHCARE
D
IRECTION
2286865, R
EVISION
14
LOGIQ™ 7 S
ERVICE
M
ANUAL
Section 6-6 - BW Printer Setting / Adjustment
6-27
Section 6-6
BW Printer Setting / Adjustment
6-6-1
Parameters for UP-D897
1.) When system started, Go to
Utility > System >Peripherals
, then hit
Printers
button in the Setup
group.
2.) Click
SONY UP-D897
in the Printer List, then right click and select "
Properties
".
3.) Select "
General
" tab, then hit "
Printing Preferences ...
"
4.) Select "
Layout
" tab, then ...
-
Set "
1920x1280
" into "Paper"
-
Check
off
into "Enlarge to Paper"
-
Set "
Bilinear
" into "Interpolation Method"
-
Set "
Landscape
" into "Orientation"
Figure 6-35 Printing Preferences
Bilinear